Tc Source Calibration Adjustment - Fluke 5560A Service Manual

Hide thumbs Also See for 5560A:
Table of Contents

Advertisement

Table 32. Calibration Steps for Inductance with LCR Meter (cont.)
Step
14
1 H
15
1.21 H
16
10 H
17
12.1 H
18
100 H
5. Push OPERATE to set the Product to Operate.
6. After the LCR meter measurement settles, input the measurement in the Enter Measurement Value
window on the Product display.
7. Continue with steps 5 to 8 until you complete all points in the sequence.
8. When a section is complete, the Product highlights the following section.

TC Source Calibration Adjustment

The Product begins the adjustment with internal steps. While internal steps are in progress, continue
as follows:
1. Set the 8588A as follows: Volts dc, Range Auto, NPLC 50, Guard OFF (internal guard).
2. Connect Type B copper mini-plug with copper wire to the meter for vdc measurements. Short the
plug at the VI AUX terminal of the Product. Allow 3 minutes for thermal dissipation.
3. Push ZERO, and then ZERO FUNC. Allow the zero function to finish.
4. When the Product shows a connect prompt, connect the 8588A for vdc measurement to the
product TC INPUT/OUTPUT as instructed on the Display with the mini-plug.
5. When the Product shows an instruction to set the Product in Operate, push OPERATE.
6. Measure with the meter and input the value on the touch screen in the Enter Measured Value field.
Note that the Product only accepts values in base units. Example: a measurement output shows as
60.00000 mV, but the Product expects the value to be input in V.
7. Follow the on-screen instructions for OPERATE/Measure/Input Value until all steps in the function
are complete.
8. When a section is complete, the Product highlights the following section.
Product Output
Calibration Adjustment
Test Frequency
10 Hz
3 Hz
3 Hz
2 Hz
2 Hz
Calibrator
73