Alarms and Messages
The pump can sound multiple alarms. For many of the alarms, you
have the option to "acknowledge" or "silence."
• Acknowledge: Clears the alarm from the screen.
• Silence: The alarm stays on the screen, but is silenced for 2 minutes.
The alarm continues until it is acknowledged or resolved.
The alarm and message types are as follows:
• System Fault Alarm: An unrecoverable
error may have occurred, such as a
hardware or software fault. With this
alarm, the amber indicator light is
continuously lit, the screen is red, and
a two-tone alarm sounds. If a system
fault alarm occurs, contact your
health care provider.
• High Priority Alarm: This alarm
always pauses or stops the pump. This
alarm is accompanied by a red screen
and persists until acknowledged by a
key press or until the condition that
triggered the alarm goes away.
If silenced, the alarm sounds again
after 2 minutes if the alarm condition
still exists.
• Medium Priority Alarm: This alarm
does not stop the pump (delivery
continues). This alarm is accompanied
by an amber screen and persists until
acknowledged by a key press or until
the condition that triggered the alarm
goes away. If silenced, the alarm
sounds again after 2 minutes if the
alarm condition still exists.
System fault alarm (red)
with error code
High priority alarm (red)
Medium priority alarm (amber)