Turning The Pump On And Off - Smiths Medical cadd-solis VIP 2120 Information Manual

Ambulatory infusion pump
Table of Contents

Advertisement

Turning the Pump On and Off

To turn on the pump,
power switch
When the pump powers on, it performs
various self-tests. Watch and listen for
the following:
• The green and amber lights flash.
• The display shows a blue and
amber rotating arc, followed by a
®
CADD
-Solis Ambulatory Infusion
System display. If you see any stripes,
or black or white pixels, or if your display appears faulty, contact
your health care provider.
• When power up completes, six beeps sound. Contact your health
care provider if you do not hear the beeps, because there may be a
problem with the audible alarms.
• If any issues are detected during power on, alarms will sound
and/or be displayed. If the alarms continue after you follow any
on-screen help, contact your health care provider.
To turn off the pump,
down?" message appears, press Yes. The pump then powers off.
When the pump is powered off:
• The display is blank.
• Keypad presses are not detected.
• Medication is not infused.
• Alarms are not sensed, and audio and visual alarms are not
displayed.
• If the AC adapter is plugged in, the blue AC power light remains
on. No other lights are lit.
• If the pump is connected to AC power and a rechargeable battery
pack is installed, the battery pack continues to charge.
16
press and hold the
.
press the power switch
Power
Switch
. When the "Power

Advertisement

Table of Contents
loading

Table of Contents