Chapter 6: Alarms; Alarm Priorities - Smiths Medical Oxi-Pulse 30 WW1030 Home Use Operation Manual

Table of Contents

Advertisement

Chapter 6: Alarms

The WW1030 is intended for continuous patient monitoring and is equipped with audible and
visual alarm indicators.
WARnInG! It is the operator's responsibility to set alarm limits appropriately for each
individual patient.
WARnInG! Always verify that the alarm volume is appropriate for the environment in
which the monitor is being used.

Alarm Priorities

Each alarm is categorized as high, medium, or low priority. In general, only the highest priority
alarm is displayed. For example, if a patient alarm (high priority) is active at the same time as a
remote alarm cable fault (medium priority), the remote alarm cable fault will not be displayed. The
following is an exception to this general rule:
During a low battery alarm (medium priority), the low battery indicator in the battery
gauge will always flash even if the other alarms are active.
nOTe: Some medium priority alarms are generated when the device detects an internal
malfunction. During these conditions, the oximeter does not monitor a patient. no
high priority alarms can occur.
nOTe: During low priority alarms conditions, the patient is not being monitored. no high
priority alarms can occur.
Home Use Operation Manual
Chapter 6: Alarms
6-1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Oxi-pulse 30Ww1030

Table of Contents