Chapter 4: Alarms

High Priority Alarms

A high priority alarm warns you when a patient's measurement matches or exceeds the high or low alarm limit for
that measurement. For example, if the low SpO
high priority alarm is triggered. During a high priority alarm:
1
2
96
The High Priority ALARM indicator flashes (red).
1
The digits for the violated alarm limit flash.
2
3
A message is displayed (subject to the display priority).
The High Priority Alarm tone sounds (if not silenced).
NOTE: The alarm actions occur for each violated alarm, even if more than one alarm is violated at the
same time.
Operation Manual SurgiVet
alarm limit is set to 92, and the patient's measured SpO
2
3
SpO2 < 98
Figure 4.1: Alarm Example
®
V3404 Plus
RR
25
/m
Chapter 4: Alarms
is 90, a
2
4-1