Turning On the Monitor

Press this key to turn
the monitor on and off.
To turn on the monitor, press the
up:
All LEDs light.
The monitor beeps briefly.
The monitor's software revisions (main, battery PIC, oximeter) are momentarily displayed.
Displays "
--HI
" if SpO
Displays "
AL OFF
Displays "
HS YES
Displays Operation Mode indicator:
Patient Record Number for Clinician Mode.
"
SLP
" for Sleep Mode.
"
" for Home Mode.
H
After a few seconds the % SpO
be shown. If not, see Chapter 9: Troubleshooting for help.
WARnInG! Verify that all leDs (light emitting diodes) on the display light up upon startup
of the device.
WARnInG! Verify that the monitor sounds a short tone upon startup of the device. If no
tone is heard, the speaker may be damaged. Do not use to monitor patients
until the monitor has been repaired.
If the WW1030 detects an error during power up, "
the display. A numeric error code will display in the SpO
for more information.
Home Use Operation Manual
Figure 5-6: Turning On the Monitor
x
key. Before using the monitor, check the following at power
high alarm limit is OFF, and not in Sleep Mode.
2
" (Alarms Off ) if in Sleep Mode.
" if High Sensitivity Mode is ON.
value, pulse rate, pulse signal strength and PI bar graphs should
2
ERR
Chapter 5: Operating Instructions
" will display in the pulse rate section of
section. See Chapter 9: Troubleshooting
2
5-7