6. Printer-specific Adjustments (GA-1211)
(11) Display the Measurement vs. Target graph of the View Measurement of the Calibrator. This graph shows the
measurement values of the calibrated gradation steps. If the bold line (tone characteristics of the target image)
and the thin line (the measured values) are almost aligned, the calibration is satisfactory. If not, retry the
calibration.
<Note>
Never click [Apply] of ColorWise Pro Tools when finishing the confirmation of the graph. If it is
clicked, the calibration results will be deleted.
(12) Print the Configuration Page and then check each field of "Date", "Media", "Measurement Method" and
"Operator" on the page to see if the calibration has been performed for each media type.
Precautions after the completion of the adjustment
Nothing in particular.