8 – Language Dictionary
MEASure:VOLTage:LOW?
FETCh:VOLTage:LOW?
Agilent 66311B/D, 66309B/D only
These queries return the Low level voltage of a voltage pulse waveform. The instrument first measures
the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse
waveform using 16 bins between the maximum and minimum data points. The bin containing the most
data points below the 50% point is the low bin. The average of all the data points in the low bin is
returned as the Low level. If no low bin contains more than 1.25% of the total number of acquired points,
then the minimum value is returned by these queries.
Query Syntax MEASure[:SCALar]:VOLTage:LOW?
Parameters None
Examples MEAS:VOLT:LOW?
Returned Parameters <NR3>
Related Commands MEAS:VOLT:HIGH?
MEASure:VOLTage:MAXimum?
FETCh:VOLTage:MAXimum?
Agilent 66311B/D, 66309B/D only
These queries return the maximum output voltage.
Query Syntax MEASure[:SCALar]:VOLTage:MAXimum?
Parameters None
Examples MEAS:VOLT:MAX?
Returned Parameters <NR3>
Related Commands MEAS:VOLT:MIN?
MEASure:VOLTage:MINimum?
FETCh:VOLTage:MINimum?
Agilent 66311B/D, 66309B/D only
These queries return the minimum output voltage..
Query Syntax MEASure[:SCALar]:VOLTage:MINimum?
Parameters None
Examples MEAS:VOLT:MIN?
Returned Parameters <NR3>
Related Commands MEAS:VOLT:MAX?
110
FETCh[:SCALar]:VOLTage:LOW?
FETCh[:SCALar]:VOLTage:MAXimum?
FETCh[:SCALar]:VOLTage:MINimum?
FETC:VOLT:LOW?
CALC:REF:LOW
FETC:VOLT:MAX?
FETC:VOLT:MIN?