TRIGger:SEQuence2:SLOPe:DVM
TRIGger:ACQuire:SLOPe:DVM
Agilent 66311D/66309D only
This command sets the slope of the DVM input signal.
POSitive
measurement triggering occurs on the rising edge.
measurement triggering occurs on the falling edge.
NEGative
measurement triggering occurs on either edge.
EITHer
Command Syntax TRIGger:SEQuence2:SLOPe:DVM<slope>
Parameters EITHer | POSitive | NEGative
*RST Value POSitive
Examples TRIG:SEQ2:SLOP:DVM POS
Query Syntax TRIGger:SEQuence2:SLOPe:DVM?
Returned Parameters <CRD>
Related Commands TRIG:SEQ2:LEV:DVM
TRIGger:SEQuence2:SLOPe:VOLTage
TRIGger:ACQuire:SLOPe:VOLTage
This command sets the slope of an internally triggered voltage measurement.
triggering occurs on the rising edge.
POSitive
NEGative
triggering occurs on the falling edge.
EITHer
triggering occurs on either edge.
Command Syntax TRIGger:SEQuence2:SLOPe:VOLTage<slope>
Parameters EITHer | POSitive | NEGative
*RST Value POSitive
Examples TRIG:SEQ2:SLOP:VOLT POS
Query Syntax TRIGger:SEQuence2:SLOPe:VOLTage?
Returned Parameters <CRD>
Related Commands TRIG:SEQ2:SLOP:CURR
TRIGger:ACQuire:SLOPe:DVM<slope>
TRIGger:ACQuire:SLOPe:DVM?
TRIGger:ACQuire:SLOPe:VOLTage<slope>
TRIGger:ACQuire:SLOPe:VOLTage?
Language Dictionary - 8
TRIG:ACQ:SLOP:DVM EITH
TRIG:ACQ:SLOP:VOLT EITH
133