CALibrate:DATA

This command enters a calibration value that you obtain by reading an external meter. You must first
select a calibration level (with CALibrate:LEVel) for the value being entered.
Command Syntax CALibrate:DATA<NRf>
Parameters <external reading>
Examples CAL:DATA 3222.3 MA
Related Commands CAL:STAT

CALibrate:DATE

Use this command to store the date that the unit was last calibrated. You can enter any ASCII string up to
10 characters.
Command Syntax CALibrate:DATE <date>
Parameters <date>
Examples CAL:DATE "3/22/99"
Query Syntax CALibrate:DATE?
Returned Parameters <SRD>

CALibrate:DVM

Agilent 66311D/66309D only
This command initiates the calibration of the DVM.
Command Syntax CALibrate:DVM
Parameters None
Examples CAL:DVM

CALibrate:LEVel

This command selects the next point in the calibration sequence. P1 is the first calibration point,
P2 is the second calibration point.
Command Syntax CALibrate:LEVel <point>
Parameters P1 | P2
Examples CAL:LEV P2

CALibrate:PASSword

This command lets you change the calibration password. A new password is automatically stored in
nonvolatile memory and does not have to be stored with CALibrate:SAVE. If the password is set to 0,
password protection is removed and the ability to enter the calibration mode is unrestricted.
Command Syntax CALibrate:PASScode<NRf>
Parameters <model number> (default)
Examples CAL:PASS 6812
Related Commands CAL:SAV
Unit A (amperes)
CAL:LEV
Language Dictionary - 8
CAL:DATA 5.000
CAL:DATE "22.3.99"