[Source:]Voltage2; [Source:]Voltage:protection; [Source:]Voltage:protection:state - Agilent Technologies 66111A User Manual

Fast transient dc source/mobile communications dc source
Table of Contents

Advertisement

8 – Language Dictionary

[SOURce:]VOLTage2

Agilent 66309B/D only
This command sets the output voltage level of the auxiliary output.
Command Syntax [SOURce:]VOLTage2[:LEVel][:IMMediate][:AMPLitude]<NRf+>
Parameters see Table A-3
Default Suffix V (volts)
*RST Value 0
Examples VOLT2 500mV
Query Syntax [SOURce:]VOLTage2[:LEVel][:IMMediate][:AMPLitude]?
Returned Parameters <NR3>

[SOURce:]VOLTage:PROTection

This command sets the overvoltage protection (OVP) level of the dc source. If the output voltage exceeds
the OVP level, then the dc source output is disabled and the Questionable Condition status register OV
bit is set (see chapter 7 about programming the status registers). An overvoltage condition can be cleared
with the OUTP:PROT:CLE command after the condition that caused the OVP trip is removed. The OVP
always trips with zero delay and is unaffected by the OUTP:PROT:DEL command. The OVP function
can be disabled using the VOLT:PROT:STAT command.
Command Syntax [SOURce:]VOLTage:PROTection[:LEVel]<NRf+>
Parameters see Table 8-3
Default Suffix V (volts)
*RST Value MAX
Examples VOLT:PROT 21
Query Syntax [SOURce:]VOLTage:PROTection[:LEVel]?
Returned Parameters <NR3>
Related Commands OUTP:PROT:CLE

[SOURce:]VOLTage:PROTection:STATe

This command enables or disables the overvoltage protection (OVP) function.
CAUTION:
Disabling the overvoltage protection function may cause excessive output voltages, such
as can occur if remote sense leads are disconnected, to damage the equipment under test.
Command Syntax [SOURce:]VOLTage:PROTection:STATe <bool>
Parameters 0 | 1 | OFF | ON
*RST Value OFF
Examples VOLT:PROT:STAT 0
Query Syntax [SOURce:]VOLTage:PROTection:STATe?
Returned Parameters <NR1>0 or 1
Related Commands OUTP:PROT:CLE
120
!set output2 voltage to 0.5V
!set voltage protect to 21V
VOLT:PROT:STAT 1
*RST
VOLT:PROT:STAT
!voltage protection is OFF
!voltage protection is ON
VOLT:PROT