Calibration Commands; Calibrate:current; Calibrate:current2; Calibrate:current:measure:lowrange - Agilent Technologies 66111A User Manual

Fast transient dc source/mobile communications dc source
Table of Contents

Advertisement

8 – Language Dictionary

Calibration Commands

Calibration commands let you enable and disable the calibration mode, change the calibration password,
calibrate current and voltage programming, and store new calibration constants in nonvolatile memory.
NOTE:
If calibration mode has not been enabled with CALibrate:STATe, programming the
calibration commands will generate an error. You must also save any changes that you made
using CALibrate:SAVE, otherwise all changes will be lost when you exit calibration mode.

CALibrate:CURRent

This command initiates the calibration of the positive dc output current as well as the high-range current
measurement circuit.
Command Syntax CALibrate:CURRent[:SOURce][:DC][:POSitive]
Parameters None
Examples CAL:STAT 1,
Related Commands CAL:CURR:NEG,

CALibrate:CURRent2

Agilent 66309B/D only
This command initiates the current calibration of output 2.
Command Syntax CALibrate:CURRent2
Parameters None
Examples CAL:CURR2
Related Commands CAL:CURR:NEG,

CALibrate:CURRent:MEASure:LOWRange

Agilent 66311B/D, 66309B/D only
This command initiates the calibration of the low-range current measurement circuit.
Command Syntax CALibrate:CURRent:MEASure[:DC]:LOWRange
Parameters None
Examples CAL:CURR:MEAS
Related Commands CAL:CURR

CALibrate:CURRent:MEASure:AC

Agilent 66311B/D, 66309B/D only
This command initiates the calibration of the high bandwidth (ac) measurement circuit.
Command Syntax CALibrate:CURRent:MEASure:AC
Parameters None
Examples CAL:CURR:MEAS:AC
100
CAL:CURR,
! enable calibration
! start current calibration
CAL:LEV,
CAL:DATA
CAL:LEV,
CAL:DATA