Status:operation; Status:operation:condition; Status:operation:enable - Agilent Technologies 66111A User Manual

Fast transient dc source/mobile communications dc source
Table of Contents

Advertisement

8 – Language Dictionary
Table 8-4. Bit Configuration of Operation Status Registers
Bit Position
15–13
Bit Name
not
used
Bit Weight
CAL = The dc source is computing new calibration constants.
WTG = The dc source is waiting for a trigger.
CV = The dc source is operating in constant voltage mode.
CV2 = Output 2 is operating in constant voltage mode. (Agilent 66309B/D only)
CC+ = The dc source is operating in constant current mode.
CC- = The dc source is operating in negative constant current mode.
CC2 = Output 2 is operating in constant current mode. (Agilent 66309B/D only)

STATus:OPERation?

This query returns the value of the Operation Event register. The Event register is a read-only register,
which holds (latches) all events that are passed by the Operation NTR and/or PTR filter. Reading the
Operation Event register clears it.
Query Syntax STATus:OPERtion[:EVENt]?
Parameters None
Returned Parameters <NR1>(Register Value)
Examples STAT:OPER?
Related Commands *CLS

STATus:OPERation:CONDition?

This query returns the value of the Operation Condition register. That is a read-only register, which holds
the real-time (unlatched) operational status of the dc source.
Query Syntax STATus:OPERation:CONDition?
Parameters None
Examples STAT:OPER:COND? STATUS:OPERATION:CONDITION?
Returned Parameters <NR1> (register value)

STATus:OPERation:ENABle

This command and its query set and read the value of the Operational Enable register. This register is a
mask for enabling specific bits from the Operation Event register to set the operation summary bit
(OPER) of the Status Byte register. This bit (bit 7) is the logical OR of all the Operatonal Event register
bits that are enabled by the Status Operation Enable register.
Command Syntax STATus:OPERation:ENABle<NRf>
Parameters 0 to 32767
Preset Value 0
Examples STAT:OPER:ENAB 1312
Query Syntax STATus:OPERation:ENABle?
Returned Parameters <NR1> (register value)
Related Commands STAT:OPER:EVEN
122
12
11
10
CC2
CC-
CC+
4096
2048
1024
STAT:OPER:NTR
9
8
7-6
CV2
CV
not
used
512
256
STAT:OPER:PTR
5
4-1
0
WTG
not
CAL
used
32
1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

66309b/d66311b/d66311b66311d66309b66309d

Table of Contents