Measure:current:high? Fetch:current:high; Measure:current:low? Fetch:current:low; Measure:current:maximum? Fetch:current: Maximum - Agilent Technologies 66111A User Manual

Fast transient dc source/mobile communications dc source
Table of Contents

Advertisement

MEASure:CURRent:HIGH?
FETCh:CURRent:HIGH?
Agilent 66311B/D, 66309B/D only
These queries return the High level current of a current pulse waveform. The instrument first measures
the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse
waveform using 16 bins between the maximum and minimum data points. The bin containing the most
data points above the 50% point is the high bin. The average of all the data points in the high bin is
returned as the High level. If no high bin contains more than 1.25% of the total number of acquired
points, then the maximum value is returned by these queries.
Query Syntax MEASure[:SCALar]:CURRent:HIGH?
Parameters None
Examples MEAS:CURR:HIGH?
Returned Parameters <NR3>
Related Commands MEAS:CURR:LOW?
MEASure:CURRent:LOW?
FETCh:CURRent:LOW?
Agilent 66311B/D, 66309B/D only
These queries return the Low level current of a current pulse waveform. The instrument first measures
the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse
waveform using 16 bins between the maximum and minimum data points. The bin containing the most
data points below the 50% point is the low bin. The average of all the data points in the low bin is
returned as the Low level. If no low bin contains more than 1.25% of the total number of acquired points,
then the minimum value is returned by these queries.
Query Syntax MEASure[:SCALar]:CURRent:LOW?
Parameters None
Examples MEAS:CURR:LOW?
Returned Parameters <NR3>
Related Commands MEAS:CURR:HIGH?
MEASure:CURRent:MAXimum?
FETCh:CURRent: MAXimum?
Agilent 66311B/D, 66309B/D only
These queries return the maximum output current.
Query Syntax MEASure[:SCALar]:CURRent:MAXimum?
Parameters None
Examples MEAS:CURR:MAX?
Returned Parameters <NR3>
Related Commands MEAS:CURR:MIN?
FETCh[:SCALar]:CURRent:HIGH?
FETCh[:SCALar]:CURRent:LOW?
FETCh[:SCALar]:CURRent:MAXimum?
Language Dictionary - 8
FETC:CURR:HIGH?
CALC:REF:HIGH
FETC:CURR:LOW?
CALC:REF:LOW
FETC:CURR:MAX?
107