B - Verification and Calibration

Performing the Calibration Procedure

NOTE:
The calibration procedure can only be performed from the front panel or using the SCPI
language commands.
Table B-1 lists the equipment required for calibration. Figure B-1 shows the test setup.
You do not have to do a complete calibration each time. If appropriate, you may calibrate only the
voltage or current and proceed to "Saving the Calibration Constants". However, the voltage or current
calibration sequence must be performed in its entirety. The following parameters may be calibrated:
voltage programming and measurement
overvoltage protection (OVP)
current programming and measurement
low range measurement
ac current measurement
internal DVM

Front Panel Calibration Menu

The Entry keypad is used for calibration functions.
Shift
Cal
Press this key to access the calibration menu.
Display
CAL ON <value>
CAL OFF
CAL:LEV <char>
CAL:DATA <value>
CAL:VOLT
CAL:VOLT2
CAL:VOLT:PROT
CAL:CURR
CAL:CURR2
CAL:CURR:MEAS:LOW
CAL:CURR:MEAS:AC
CAL:DVM
CAL:SAVE
DATE <char>
CAL:PASS <value>
1
not valid for Agilent 66111A
Notes:
value = a numeric value
char = a character string parameter
p
Use
È
Use
Æ
Use
154
Command Function
Turns calibration mode on when the correct password
value is entered.
Turns calibration mode off
Advance to next step in sequence (P1 or P2).
Enter an external calibration measurement.
Begin voltage calibration sequence
Begin output 2 voltage calibration sequence
Begin voltage protection calibration
Begin high range current calibration sequence
Begin output 2 current calibration sequence
Begin low range current measurement calibration
Begin ac current calibration sequence
Begin DVM calibration sequence
Saves the calibration constants in non-volatile memory.
Displays the calibration date. (0 if none is supplied)
Set new calibration password.
q
and
to scroll through the menu commands.
É
and
to scroll through the menu parameters.
Ç
and
to select a digit in a numeric entry field.