Basic - Agilent Technologies 66111A User Manual

Fast transient dc source/mobile communications dc source
Table of Contents

Advertisement

D – Example Programs

BASIC

The BASIC for controllers and BASIC for Windows both provide access to GPIB functions at the
operating system level. This makes it unnecessary to have the header files required in front of DOS
applications programs. Also, you do not have to be concerned about controller "hangups" as long as your
program includes a timeout statement. Because the dc source can be programmed to generate SRQ on
errors, your program can use an SRQ service routine for decoding detected errors. The detectable errors
are listed in Appendix C.
Example 2. Controller Using BASIC
1000
!Dc source at stand-alone address = 706
1005
OPTION BASE 1
1010
DIM Codes$[80],Response$[80],Mode$[32]
1015
!
1020
!Program dc source to CV mode with following voltage and current
1025
OUTPUT 706;"VOLTAGE MAX;CURRENT MAX"
1030
!
1035
!Query dc source outputs and print to screen
1040
OUTPUT 706;"MEASURE:VOLTAGE?;CURRENT?"
1045
ENTER 706;Vout,Iout
1050
PRINT "The output levels are ";Vout;" Volts and ";Iout" Amps"
1055
!
1060
!Program current triggered level to a value insufficient to maintain
1065
!supply within its CV operating characteristic
1070
OUTPUT 706;"CURR:TRIG MIN"
1075
!
1080
!Set operation status mask to detect mode change from CV to CC
1085
OUTPUT 706;"STAT:OPER:ENAB 1024;PTR 1024"
1090
!
1095
!Enable Status Byte OPER summary bit
1100
OUTPUT 706;"*SRE 128"
1105
!
1110
!Arm trigger circuit and send trigger to dc source
1115
OUTPUT 706;"INITIATE:SEQUENCE1;TRIGGER"
1130
!Poll for interrupt caused by change to CC mode and print to screen
1135
Response=SPOLL(706)
1140
IF NOT BIT (Response,7) THEN GOTO 1130
1145
OUTPUT 706;"STAT:OPER:EVEN?"
1150
ENTER 706;Oevent
1155
IF BIT(Oevent,10) THEN PRINT "Supply switched to CC mode."
1160
!
1165
!Clear status
1170
OUTPUT 706;"*CLS"
1175
!
1180
!Disable output and save present state in location 2
1185
OUTPUT 706;"OUTPUT OFF;*SAV 2"
1190
END
168
!Query output levels
!No OPER event to report
!Query status operation register
!Read back event bit

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

66309b/d66311b/d66311b66311d66309b66309d

Table of Contents