Calibration Commands

Calibration commands let you enable and disable the calibration mode, change the calibration password,
calibrate current and voltage programming, and store new calibration constants in nonvolatile memory.
NOTE:
If calibration mode has not been enabled with CALibrate:STATe, programming the
calibration commands will generate an error. You must also save any changes that you made
using CALibrate:SAVE, otherwise all changes will be lost when you exit calibration mode.
CALibrate:CURRent
This command initiates the calibration of the positive dc output current as well as the high-range current
measurement circuit.
Command Syntax CALibrate:CURRent[:SOURce][:DC][:POSitive]
Parameters None
Examples CAL:STAT 1,
Related Commands CAL:CURR:NEG,
CALibrate:CURRent2
Agilent 66319B/D only
This command initiates the current calibration of output 2.
Command Syntax CALibrate:CURRent2
Parameters None
Examples CAL:CURR2
Related Commands CAL:CURR:NEG,
CALibrate:CURRent:MEASure:R3
This command initiates the calibration of the middle-range current measurement circuit.
Command Syntax CALibrate:CURRent:MEASure:R3
Parameters None
Examples CAL:CURR:MEAS:R3
Related Commands CAL:CURR
CALibrate:CURRent:MEASure:LOWRange
This command initiates the calibration of the low-range current measurement circuit.
Command Syntax CALibrate:CURRent:MEASure[:DC]:LOWRange
Parameters None
Examples CAL:CURR:MEAS
Related Commands CAL:CURR
94
CAL:CURR,
! enable calibration
! start current calibration
CAL:LEV,
CAL:DATA
CAL:LEV,
CAL:DATA