SECTION 14—COOLING FAN
SECTION 14—COOLING FAN
NOTE: Refer to Recycling Information following local governing ordinances and recycling plans
regarding disposal of the concentrator or components.

Replacing Cooling Fan

To prevent electrical shock, ALWAYS disconnect concentrator from electrical
outlet before servicing.
1. Turn power Off (O) and unplug the unit.
2. Remove cabinet. Refer to Cabinet on page 25.
3. Remove control panel. Refer to Removing Control Panel on page 50.
4. Remove compressor assembly. Refer to Compressor on page 35.
5. Disconnect and remove spade connectors from fan terminals on back side of cooling
fan (not shown).
6. Remove four mounting screws securing the cooling fan to the sound box.
7. Remove existing cooling fan.
Fan MUST be positioned properly so air from the fan blows down onto the
compressor (see air flow arrow on back of fan and make sure the arrow is pointing
down) or damage to the unit will occur.
8. Install new cooling fan.
NOTE: Run the self-threading mounting screws through the fan housing holes to cut the threads
and remove them before mounting the fan for ease of installation.
9. Reinstall compressor. Refer to Compressor on page 35.
10. Reinstall control panel. Refer to Removing Control Panel on page 50 and reverse all
steps.
11. Run unit and inspect for leaks. Refer to Leak Test on page 71.
12. If no leaks detected, turn power Off (O) and unplug the unit.
13. Reinstall cabinet. Refer to Removing Cabinet on page 25.
®