3. Remove hoist from track. Refer to 3.2 Dismounting the
ceiling hoist, page 10.
4. Remove the cover. Refer to 3.4 Removing the hoist
cover, page 13.
5. Reset the emergency stop by pressing the reset button.
6. Check that the strap A is completely unwound as
shown in the image. If not, wind the straps out using
the UP/DOWN button on the hand control.
7. Align the strap retaining bolts B with the aperture in the
chassis using the UP/DOWN button on the hand control.
8. Pull the red emergency strap again.
9. Remove strap bolts, using a 10 mm socket key and a
5 mm hex/Allen key.
10. Remove the strap through guide cylinder.
14
11. Insert a new strap through guide cylinder and assemble
using the strap retaining bolts (Torque: 4 Nm). Make
sure that the new strap is fed over the slack strap
switch wire.
12. Re-mount the cover, re-assemble the hoist to the track
and reset the emergency stop.

3.6 Replacing the battery

CAUTION!
Static electricity can damage the PCBs inside the
hoist.
Before working in the inside of the hoist:
– Pull the red emergency strap to turn off the
hoist.
– Remove the hoist from the track system. (Refer
to 3.2 Dismounting the ceiling hoist, page 10)
– Remove the cover from the hoist. (Refer to 3.4
Removing the hoist cover, page 13)
– An anti-static mat must be used when working
inside the ceiling hoist.