Figure 9 User Calibration Cable For 2-Channel Oscilloscope - Agilent Technologies InfiniiVision 7000 Series Service Manual

Hide thumbs Also See for InfiniiVision 7000 Series:
Table of Contents

Advertisement

3
Calibrating and Adjusting
To Channel 1
62
Use 50
RG58AU or equivalent BNC cables when
performing User Cal.
a For a 2- channel oscilloscope, connect a BNC tee to the
equal length cables. Then connect a BNC(f)- to- BNC(f)
(also called a barrel connector) to the tee as shown
below.
User Calibration cable for 2-channel oscilloscope
Figure 9
b For a 4- channel oscilloscope, connect BNC tees to the
equal- length cables as shown below. Then connect a
Longer cable
To Channel 2
7000 Series Oscilloscopes Service Guide
to TRIG OUT