Agilent 7000/7010
Series Triple Quadrupole
Operating Manual
Agilent Technologies

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 7010 Series

  • Page 1 Agilent 7000/7010 Series Triple Quadrupole Operating Manual Agilent Technologies...
  • Page 2 Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 3: Table Of Contents

    Tightening Column Nut To Install a Capillary Column in the GC/MS Interface Using a Standard Column Nut To Install the Interface Tip Seal The GC/MS Interface for a 7890 Series GC To Calibrate the Column 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 4 Front Panel Instrument Status LED Before You Turn On the MS Pumping Down Controlling Temperatures Controlling Column Flow Controlling CC Flow Controlling JetClean Hydrogen Flow Venting the MS High Vacuum Pressure in EI Mode 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 5 CI Autotune The CI Flow Control Module To Operate the Reagent Gas Flow Control Module To Set a Reagent Gas Flow To Perform a CI Autotune To Pumpdown and Switch from EI to CI Mode 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 6 To Remove the EI HES Filaments To Install the EI HES Filaments To Close the Front Analyzer Chamber To Remove Left Rear Cover To Open the Rear Analyzer Chamber To Replace the EM Horn 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 7 To Install the CI/EI XTR Source Radiator To Switch from the CI Source to the EI HES To Install a CI Source Filament To Remove the CI Source Radiator from the 7010 Series TQ To Disassemble the CI Source To Clean the CI Source...
  • Page 8 To Remove the CI Source Filament To Install a CI Source Filament 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 9 This manual contains information for operating and maintaining the Agilent 7000/7010 Series Triple Quad (TQ) Mass Spectrometer (MS) system. “Introduction” Chapter 1 describes general information about the 7000/7010 Series TQ GC/MS, including a hardware description, general safety warnings, and hydrogen safety information.
  • Page 10 This can be installed by anyone who has authority to copy information onto the receiving computer. See the Agilent 7000/7010 Series TQ GC/MS System Quick Start document for information on finding and installing the documentation located on this USB and DVDs.
  • Page 11 Cleaning/Recycling the Product Accidental Liquid Spill Moving or Storing the MS This section provides general information about the 7000/7010 Series Triple Quad (TQ) Gas Chromatograph (GC) / Mass Spectrometer (MS), including a hardware description, general safety warnings, and hydrogen safety information.
  • Page 12: Introduction

    High-energy dynode (refers to detector and its power supply) High Efficiency Source Inside diameter Inert Standard EI source constructed of inert materials Inert+ MS model equipped with an EI XTR source Local Area Network 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 13 Perfluoro-5,8-dimethyl-3,6,9-trioxydodecane (calibrant) PFTBA Perfluorotributylamine (calibrant) Quad Quadrupole mass filter Radio frequency RFPA Radio frequency power amplifier Torr Unit of pressure, 1 mm Hg Triple Quad (Quadrupole) Turbo Split flow turbomolecular vacuum pump EI Extractor source 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 14: The 7000/7010 Series Tq Gc/Ms

    Introduction The 7000/7010 Series TQ GC/MS The 7000/7010 Series TQ GC/MS is a standalone capillary GC detector for use with the Agilent 7890 Series GC. (See Figure 1 on page 16.) The TQ GC/MS features: • One split flow turbomolecular vacuum pump •...
  • Page 15: 7000/7010 Ci Ms System

    (high vacuum) in the vacuum manifold and at the turbomolecular vacuum pump discharge. 7000/7010 CI MS System In this manual, the term “CI MS” refers to the 7000/7010 Series TQ GC/MS CI source system. It also applies, unless otherwise specified, to the mass flow controller (MFC) for these instruments.
  • Page 16: Tq Gc/Ms Hardware Description

    Introduction TQ GC/MS Hardware Description 7000/7010 Series 7890B TQ MS series GC Front panel MS power Drawer switch GC power switch Figure 1 7000/7010 Series TQ GC/MS 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 17: Important Safety Warnings

    The printed circuit boards in the MS can be damaged by electrostatic discharge (ESD). Do not touch any of the boards unless it is absolutely necessary. If you must handle them, wear a grounded wrist strap and take other antistatic precautions. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 18 WA R N I N G under, over, or around the foreline (roughing) pump constitute a fire hazard. Keep the pan clean, but do not leave absorbent material such as paper towels in it. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 19: Hydrogen Safety

    • Accumulation of hydrogen in the GC oven and subsequent combustion. See your GC documentation and the label on the top edge of the GC oven door. • Accumulation of hydrogen in the MS and subsequent combustion 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 20 If the power fails, both the GC and MS shut down. The flow of carrier gas, however, is not necessarily shut down. As described previously, in some GCs a power failure may cause the carrier gas flow to be set to maximum. As a result, hydrogen may accumulate in the MS. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 21 • Do not allow hydrogen from a high pressure cylinder to vent directly to atmosphere (danger of self-ignition). • Use a hydrogen generator instead of bottled hydrogen. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 22: Safety And Regulatory Certifications

    Always ensure the vent line is vented into a fume hood. Safety and Regulatory Certifications The 7000/7010 Series TQ GC/MS conforms to the following safety standards: • Canadian Standards Association (CSA): CAN/CSA-C222 No. 61010-1-04 • CSA/Nationally Recognized Test Laboratory (NRTL): UL 61010–1 •...
  • Page 23 이 기기는 업무용 (A 급 ) 전자파적합기기로서 판 매자 또는 사용자는 이 점을 주 의하시기 바라 며 , 가정외의 지역에서 사용하는 것을 목적으 로 합니다 . Information The Agilent Technologies 7000/7010 Series TQ GC/MS meets the following IEC classifications: Equipment Class I, Laboratory Equipment, Installation Category II, and Pollution Degree 2.
  • Page 24 Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
  • Page 25 • Ensure that all peripheral devices are also certified. • Ensure that appropriate cables are used to connect the device to peripheral equipment. • Consult your equipment dealer, Agilent Technologies, or an experienced technician for assistance. Changes or modifications not expressly approved by Agilent Technologies could void the user’s authority to operate the equipment.
  • Page 26: Intended Use

    The MS must remain upright at all times; this requires special caution when moving. The MS should not be left vented to atmosphere for long periods. (See “To Set the Interface Temperature from the 7890 GC” on page 101.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 27 To Install the Interface Tip Seal The GC/MS Interface for a 7890 Series GC To Calibrate the Column This chapter shows you how to select, install, and condition a GC column, which must be performed before you operate your MS. Agilent Technologies...
  • Page 28: Installing 7890 Gc Columns

    0.4 mL/min Total gas flow into the MS = column flow + CC gas flow + reagent gas flow (if applicable) + JetClean H flow (if applicable) † Expect degradation of spectral performance and sensitivity. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 29 Tips and hints • The column installation procedure for the 7000/7010 Series TQ GC/MS is different from that for previous MSs. Using the procedure from another instrument may not work and may damage the column or the MS.
  • Page 30: To Install A Capillary Column In A Split/Splitless Inlet

    WA R N I N G sure they are cool. Cool the oven to room temperature. Always wear clean gloves while handling any parts that go inside the GC or analyzer C A U T I O N chambers. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 31 6 mm past the end of the ferrule. (See Figure 3 on page 32.) Insert the column in the inlet. Slide the nut up the column to the inlet base and finger-tighten the nut. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 32 13 Verify flow by submerging the free end of the column in isopropanol. Look for bubbles. Insulation cup Reducing nut Capillary column 4 to 6 mm Ferrule (inside nut) Inlet column nut Septum Figure 3 Installing a capillary column for a split/splitless inlet 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 33: To Condition A Capillary Column

    Attach the column to the detector. See also For more information about installing a capillary column, refer to Optimizing Splitless Injections on Your GC for High Performance MS Analysis, Agilent Technologies publication number 5988-9944EN. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 34: To Install A Capillary Column In The Gc/Ms Interface Using The Self Tightening Column Nut

    0.40 mm id, for 0.25 mm id columns (5181-3323) • 0.5 mm id, for 0.32 mm id columns (5062-3514) • 0.8 mm id, for 0.53 mm id columns (5062-3512) • Septum (may be old, used inlet septum) • Safety glasses 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 35 Inspect the end for jagged edges or burrs. If the break is not clean and even, repeat steps 6 and 7. 10 Adjust the column so it extends this specified distance from the end of the transfer line. (See Figure 4 Figure 5 on page 36.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 36 Installing a capillary column in the GC/MS interface for an EI XTR or a CI source column extends 4-5 mm Transfer line Tip holder assembly Figure 5 Installing a capillary column in the GC/MS interface for an EI HES 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 37 14 Install the tip seal and spring on the GC/MS interface. Align, gently slide, and screw the tip seal into the tip seal holder. (See Figure 4 Figure 5 on page 36.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 38 16 You can align the ion source and interface tip seal by wiggling the side plate on its hinge. If the door still will not close, contact your Agilent Technologies service representative. 17 Close the front analyzer chamber. (See “To Close the Front Analyzer Chamber” on page 169.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 39: To Install A Capillary Column In The Gc/Ms Interface Using A Standard Column Nut

    0.5 mm id, for 0.32 mm id columns (5062-3506) • 0.8 mm id, for 0.53 mm id columns (5062-3512) • Septum (may be old, used inlet septum) • Safety glasses • Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 40 Inspect the end for jagged edges or burrs. If the break is not clean and even, repeat steps 6 and 7. 10 Adjust the column so it extends this specified distance from the end of the transfer line. (See Figure 4 Figure 5 on page 36.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 41 16 You can align the ion source and interface tip seal by wiggling the side plate on its hinge. If the door still will not close, contact your Agilent Technologies service representative. 17 Close the front analyzer chamber. (See “To Close the Front Analyzer Chamber” on page 169.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 42: To Install The Interface Tip Seal

    Thread the knurled tip seal retainer into the tip seal hold finger tight. Interface tip seal Knurled tip seal retainer Tip seal holder Figure 7 Interface tip seal 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 43 5 You can align the analyzer and interface by wiggling the side plate on its hinge. If the analyzer still will not close, contact your Agilent Technologies service representative.
  • Page 44: The Gc/Ms Interface For A 7890 Series Gc

    Column guide tube Tip seal assembly Column Ionization chamber Ion source GC oven Analyzer chamber Optional entry for CI reagent gas or JetClean system hydrogen Heater/Sensor assembly Figure 8 The GC/MS interface for a 7890B GC 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 45 C A U T I O N GC oven, or the inlet. The GC/MS interface operates at high temperatures. If you touch it when it is hot, it WA R N I N G will burn you. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 46: To Calibrate The Column

    Change any parameters that are different than those used in your method. (See Figure 9 on page 47.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 47 Installing 7890 GC Columns Figure 9 Calculate Column Length dialog 11 When the new column length appears, click OK to save the changes. 12 Click OK on the Calibrate Columns dialog to save the calibration. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 48 Installing 7890 GC Columns 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 49 GC/MS Interface into the ion source, and maintain the column guard chip. If you are using the Agilent Intuvo 9000 Gas Chromatograph with your MS, Chemical Ionization (CI) and the JetClean system are not currently supported. Agilent Technologies...
  • Page 50: Intuvo 9000 Gc Column Maintenance

    † Maximum gas flow, mL/min Maximum column id 0.53 mm (30 m length) Total gas flow into the MS = column flow + CC gas flow † Expect degradation of spectral performance and sensitivity 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 51: Conditioning Intuvo Columns

    MS power. If the foreline pump is off, hydrogen will accumulate in the MS and an explosion may occur. Read “Hydrogen Safety” on page 19 before operating the MS with hydrogen gas. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 52: Handling The Intuvo 9000Gc Column And Bus Components

    • Avoid scratching or deforming click and run sealing surfaces. • If you need to clean a sealing surface, use clean, compressed air. • Use a new gasket each time you install a column or Intuvo chip. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 53: To Replace An Intuvo Gc Column

    “To Vent the MS” page 99.) If you are using hydrogen or other flammable gas as a carrier gas, close the manual gas supply valve to the instrument before turning off the MS power. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 54 Open the bus door and remove it by lifting the door vertically off its hinge pins. Lower the oven door. Using the Intuvo torque driver, rotate all four column clamps off the column's retainer ring. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 55 Using the Intuvo torque driver, remove the two compression bolts that seal the column's click and run connectors to the bus and store for later use. 10 Remove and store the column for later use in accordance with the column manufacturer's recommendations. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 56 Over-tightening can damage the flow path, strip the fittings, and cause leaks. 18 Tighten the compression bolts until you hear a click from the Intuvo torque driver. 19 Close the column door. 20 Install the bus door. 21 Close the GC front door. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 57: To Replace An Intuvo 9000 Gc Gasket

    Check that the gasket's round lobes fit flat against the bus click and run fitting. The new gasket is ready for the chip or column attachment. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 58: To Install A Column Guard Or Jumper Chip

    Inlets > Guard Chip > Prepare for Maintenance >Replace Liner and Guard Chip > Start Maintenance. This procedure cools the inlet, detector, column, Guard Chip, and other components in the flow path heated zones to < 40 °C and configures the GC. Follow the on-screen GC prompts. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 59 Open the bus door and remove it by lifting the door vertically off its hinge pins. Pull the compression bolt access plate out to allow the torque driver to access the Guard chip compression bolt. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 60 16 Finger tighten the compression bolt. 17 Raise the Guard Chip heater. 18 Finger tighten the Guard Chip screw in the inlet base. 19 Tighten the Guard Chip screw in the inlet base with a 7/16-inch open end wrench. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 61 21 Install the inlet cover. 22 Install the bus door on its hinges and close it. 23 Close the GC front door. 24 If removed, install the ALS injector. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 62: To Replace The 9000 Gc/Ms Tail

    9000 GC/MS Tail, and other components in the flow path heated zones to < 40 °C. If you are using hydrogen or other flammable gas as a carrier gas, close the manual gas supply valve to the instrument before turning off the MS power. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 63 MS tail heater assembly clamp Compression bolts Figure 14 MS tail heater assembly clamp closed 10 Open the MS tail heater assembly by loosening the thumb screw at the top of the clamp and rotating the clamp down. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 64 14 Carefully slide the 9000 GC/MS Tail into the GC/MS interface, and gently place the click and run connection into the bus connection. 15 Check that the 9000 GC/MS Tail click and run fitting sits flat against the gasket in the bus connection. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 65 26 You can align the ion source and interface tip seal by wiggling the side plate on its hinge. If the door still will not close, contact your Agilent Technologies service representative. 27 Close the front analyzer chamber. (See “To Close the Front Analyzer Chamber” on page 169.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 66: To Condition An Intuvo Capillary Column

    Use MassHunter to run an air and water check. Proceed to the next step if the air and water are within limits. Program the column temperature to increase from 120 °C to the maximum temperature limit for the column at a rate of 10 to 15 °C/min. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 67 Never exceed the maximum column temperature, in the GC/MS interface, the GC oven, C A U T I O N or the inlet. Hold at the maximum temperature for 30 minutes. The column is conditioned and ready to be used with your method. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 68: To Install The Interface Tip Seal

    Thread the knurled tip seal retainer into the tip seal hold finger tight. Interface tip seal Knurled tip seal retainer Tip seal holder Figure 17 Interface tip seal 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 69: The Gc/Ms Interface For A 9000 Series Gc

    5 You can align the analyzer and interface by wiggling the side plate on its hinge. If the analyzer still will not close, contact your Agilent Technologies service representative.
  • Page 70 HES. The 9000 GC/MS Tail should be operated in the 250  to 350 C range. Subject to that restriction, the transfer line temperature should be slightly higher than the column temperature. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 71 To Vent the MS To Set the Interface Temperature from the 7890 GC To Save a Method on the 7890 GC This chapter explains how to perform some routine operating procedures for the 7000/7010 Series TQ GC/MS in EI mode. Agilent Technologies...
  • Page 72: Operating In Ei Mode

    Your 7000 TQ operates in EI mode using the EI XTR Ion Source Assembly, and your 7010 TQ operates in EI mode using the EI HES Source Assembly, which provides higher sensitivity when ionizing the sample. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 73: Set The Ip Address And View Tq Parameters With The Wui

    WUI Main Menu page. (See Figure 19.)  If the IP address was never configured, the factory defaults for the IP address, gateway, and subnet are 192.168.254.12, 0.0.0.0, and 255.255.255.0 respectively. Figure 19 WUI main menu 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 74 Agilent recommends attaching a label to the TQ recording these NetConfig parameters. Close the browser page when you are done. Run the Agilent GC/MS Instrument Configuration program. To configure the TQ for DCOMM you must first configure the GC before configuring the 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 75 Press the MS On/Off switch to start the instrument. Enter the instrument IP address into a PC web browser URL to display the WUI Main Menu page. (See Figure 21.) Figure 21 WUI main menu 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 76 Click Instrument Status to display the TQ turbo pump speed, zone temperatures, high vacuum pressure, TQ faults, net configuration parameters, and smart card firmware version. (See Figure 22.) Figure 22 WUI instrument status When done, close the browser. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 77: Front Panel Instrument Status Led

    JetClean Acquire & Clean operation Blinking magenta JetClean Clean Only operation Solid magenta Ready and not connected to DS Solid blue Start up (prior to FW load) Blinking red (<2 sec) Fault Solid red 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 78: Before You Turn On The Ms

    JetClean system supply shutoff valves must be closed before turning off the MS power. If the foreline pump is off, hydrogen will accumulate in the MS and an explosion may occur. Read “Hydrogen Safety” on page 19 before operating the MS with hydrogen gas. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 79: Pumping Down

    Normally, the GC/MS interface heater is powered and controlled by the Thermal Aux #2 heated zone of the GC. The GC/MS interface temperature can be set and monitored from the data system or from the GC. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 80: Controlling Column Flow

    GC/MS interface where it then enters the ion source body. The flow rate is controlled by the MassHunter GC/MS Acquisition software. (See “To Set the Mode for the Optional JetClean System” on page 89.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 81: Venting The Ms

    Do not vent while the turbo pump is still spinning at more than 50%. Do not exceed the maximum recommended total gas flow. (See Table 3 on page 28.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 82: High Vacuum Pressure In Ei Mode

    1.34 * 10 If the pressure is consistently higher than those listed, refer to the online help in the Agilent MassHunter GC/MS Acquisition software for information on troubleshooting air leaks and other vacuum problems. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 83: To Set Monitors For Ms Temperature And Vacuum Status

    The available monitors may change due to the configuration of the GC and In the Available Monitors column, select MS High Vac and click Add to move the selection to the Selected Monitors column. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 84 10 Select Window > Arrange Plots and Monitors, or click and drag each monitor to the desired position. See Figure 24 for one way of arranging the monitors. Figure 24 Arranging monitors 11 To make the new settings part of the method, select Save from the Method menu. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 85: To Set The Ms Analyzer's Temperatures

    Select the Files and Reports tab, then click Save to save the tune file with these changes. Table 7 Recommended temperature settings Zone EI operation °C MS Source 250 - 280 °C MS Quad 1 °C MS Quad 2 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 86 The software will not allow you to exceed 200 °C for the quadrupole or 350 °C for the WA R N I N G source. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 87: To Set The Gc/Ms Interface Temperature From Masshunter

    Click Apply to download setpoints or click OK to download setpoints, and close the window. To make the new settings part of the method, select Save from the Method menu. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 88: To Configure Cc Gas

    Configure CC gas From the Collision Cell EPC >Quench Gas drop-down menu, select the quench gas. From the Collision Cell/EPC > Collision Gas drop-down menu, select the collision gas. Click OK to save the configuration. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 89: To Set The Mode For The Optional Jetclean System

    Click Source Cleaning from the Triple Quadrupole MS Method Editor navigation panel. Select a mode from the Operation dropdown menu. (See Figure 28.) Figure 28 TQ MS Method Editor Source Cleaning panel For detailed information on the JetClean system see the JetClean Operating manual. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 90: To Set The Jetclean Parameters For The Clean Only Mode

    Select Source Cleaning from the Triple Quadrupole MS Method Editor navigation panel. Select Clean Only from the Operation dropdown menu. (See Figure 29.) Figure 29 JetClean set to Clean Only mode Enter the parameters supplied by your method developer. Save the method. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 91: To Set The Cc Gas Flow Rates

    Click Apply to download the setpoints or OK to download the setpoints and close the window. To make the new settings part of the method, select Save from the Method menu. Figure 30 Setting CC gas flow rate 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 92: To Autotune The Ms For Ei Mode

    Do not select this item if you want to review the autotune report before saving the newly generated tune parameters. Select the Print autotune report check box to automatically print a tune report. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 93 Save tune file when done check box, save the autotune by clicking the Files and Report tab, then click Save. Refer to the manuals or online help provided with your Agilent MassHunter GC/MS Acquisition software for additional information about tuning. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 94: To Open Left Side Panel For Access To The Analyzer Chambers

    (See Figure 32.)   Procedure Pull gently on the front left window, and allow the left panel to slide forward and down. Left panel Figure 32 The left panel 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 95: To Pumpdown The Ms

    JetClean system is installed. Refer to the JetClean Operating manual for warnings regarding hydrogen safety when the hydrogen shutoff valve is open. Verify that the vent valve is open by turning it clockwise until closed. (See Figure 33 on page 96.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 96 The foreline pump will make a gurgling noise. This noise should stop within a minute. If the noise continues, there is a large air leak in your system, probably at the sideplate seal, the interface column nut, or the vent valve. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 97 14 The Pumpdown dialog displays. (See Figure 34.) You are prompted to make sure the vent valve is closed. 15 Click OK to close the Pumpdown manual step dialog and start the Pumpdown. Figure 34 Pumpdown status 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 98 92 or “To Perform a CI Autotune” on page 116.) 19 If using the JetClean system, prepare the hydrogen system for use. Refer to the JetClean Operating manual for warnings regarding hydrogen safety. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 99: To Vent The Ms

    GC/MS interface heater and the GC oven temperatures to ambient (room temperature). Turn off the MS by pressing the power switch. (See Figure 1 on page 16.) Unplug the MS power cord. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 100 Allow the analyzers to cool to near room temperature before touching them. WA R N I N G Always wear clean gloves while handling any parts that go inside the analyzer C A U T I O N chambers. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 101: To Set The Interface Temperature From The 7890 Gc

    You can also upload the GC method to the MassHunter Workstation to save the new setpoints made at the GC keypad. When a new method is loaded, all the setpoints in the new method will overwrite those currently in the GC. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 102: To Save A Method On The 7890 Gc

    A message is displayed if a method with the number you selected already exists. Press [On/Yes] to replace the existing method or [Off/No] to return to the stored methods list without saving the method. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 103 To Pumpdown and Switch from EI to CI Mode This chapter provides information and instructions for operating the 7000/7010 Series TQ GC/MS system in CI mode. Most of the information in the preceding chapter is also relevant. Most of the material is related to methane CI, but one section discusses the use of other reagent gases.
  • Page 104: Operating In Ci Mode

    • Ensure the reagent gas plumbing has no air leaks. This is determined in PCI mode, checking for m/z 32 after the methane pretune. • Ensure the reagent gas inlet line(s) are equipped with gas purifiers (not applicable for ammonia). 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 105: Operating The Ci Ms

    Depending upon the application, use the following reagent gas flowrates during system startup: • PCI mode set reagent gas flow to 20 (1 mL/min) • NCI mode set reagent gas flow to 40 (2 mL/min) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 106: High Vacuum Pressure In Ci Mode

    × × 2.10 7.31 1.71 4.30 10 –1 –5 10 –1 –6 × × × × 2.18 7.39 1.80 4.76 10 –1 –5 10 –1 –6 × × × × 2.25 7.43 1.88 5.18 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 107: Other Reagent Gases

    Other Reagent Gases This section describes the use of isobutane or ammonia as the reagent gas. You should be familiar with operating the CI-equipped 7000/7010 Series TQ GC/MS with methane reagent gas before attempting to use other reagent gases. Do not use nitrous oxide as a reagent gas. It radically shortens the life span of the C A U T I O N filament.
  • Page 108 Ammonia tends to break down vacuum pump fluids and seals. Ammonia CI makes more frequent vacuum system maintenance necessary. Refer to the 7000/7010 Series Triple Quad GC/MS Troubleshooting and Maintenance Manual. When running ammonia for 5 or more hours a day, the foreline pump must be ballasted C A U T I O N (flushed with air) for at least 1 hour a day to minimize damage to pump seals.
  • Page 109: Ci Autotune

    130 V 130 V 130 V 130 V 130 V 130 V Entrance lens offset 20 V 20 V 20 V 20 V 20 V 20 V EM volts 1200 1400 1200 1400 1200 1400 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 110 150 °C Rear quad temp 150 °C 150 °C 150 °C 150 °C 150 °C 150 °C Interface temp 280 °C 280 °C 280 °C 280 °C 280 °C 280 °C Autotune N/A Not available 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 111: The Ci Flow Control Module

    Gas A (methane) supply Shutoff Gas A valve select valve Gas B select valve GC/MS Calibration Gas B interface valve (other) supply Restrictor Calibration vial GC column Figure 36 Reagent gas flow control module schematic 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 112 Closed       setpoint setpoint 100% 100% 100% Shutoff valve Open Open Open Open Open Closed The Open and Closed states are shown in the monitors as Opened and Closed. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 113: To Operate The Reagent Gas Flow Control Module

    In Instrument Control panel, select the MS Tune icon to display the Triple Quadrupole MS Tune window. Select the Manual Tune tab, then select the Ion Source tab to display the ion source parameters. (See Figure 37.) Figure 37 CI flow control 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 114 Pumpout - Closes both reagent gas valves for 4 minutes, and evacuates the system of reagent gases. At the end of the pumpout time, the selected reagent gas valve opens. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 115: To Set A Reagent Gas Flow

    The reagent gas is flowing into the ion source at the rate displayed next to the setpoint. Click the Files and Reports tab, then click Save to save your changes to the currently loaded tune file. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 116: To Perform A Ci Autotune

    Gas section, select CI Gas A or CI Gas B as the reagent gas, and then enter a CI Gas Flow rate of 20% for a PCI source or 40% for an NCI source. Click on the Autotune tab to return to Autotune. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 117 12 Review the Tune Report. If the results are acceptable and you did not select the Save tune file when done check box, save the autotune by clicking the Files and Report tab, then click Save. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 118: To Pumpdown And Switch From Ei To Ci Mode

    PCI mode before switching to NCI. In Instrument Control view, select Instrument > Gas Controller Configuration to display the Gas Controller Configuration dialog. (See Figure 38.) Figure 38 Gas controller configuration for CI 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 119 PCI or NCI source and reagent gas setting. 12 Select the Tune from default settings check box because you have changed the ion source. 13 Select the Print autotune report check box to automatically print a tune report. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 120 • Maximum emission current: Optimum emission current maximum for NCI is very compound-specific and must be selected empirically. Optimum emission current for pesticides, for example, may be about 200 µA. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 121 To Remove the EI HES Filaments To Install the EI HES Filaments To Close the Front Analyzer Chamber To Remove Left Rear Cover To Open the Rear Analyzer Chamber To Replace the EM Horn To Close the Rear Analyzer Chamber Agilent Technologies...
  • Page 122: General Maintenance

    Leak check the system As needed Or as needed. † Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be lubricated. Lubricating other seals can interfere with their correct function. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 123 Some procedures in this chapter require access to the inside of the MS while the power switch is on. Do not remove any of the electronics safety covers in any of these procedures. To reduce the risk of electric shock, follow the procedures carefully. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 124 An oil trap is supplied with the standard foreline pump. This trap stops only pump oil droplets. It does not trap any other chemicals. If you are using toxic solvents or analyzing toxic chemicals, do not use this oil trap. Install a hose to 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 125 Always purge the MS with methane after using ammonia. Install the ammonia so the tank is in an upright position. This will help prevent liquid ammonia from getting into the flow module. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 126 Antistatic precautions are not 100% effective. Handle electronic circuit boards as little as possible, and then only by the edges. Never touch components, exposed traces, or pins on connectors and cables. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 127: Maintaining The Vacuum System

    More information is available If you need more information about the locations or functions of vacuum system components, refer the Agilent 7000/7010 Series Triple Quad GC/MS Troubleshooting and Maintenance Manual. Most of the procedures in this chapter are illustrated with video clips on the Agilent GC/MS Hardware User Information &...
  • Page 128: Maintaining The Analyzer

    • Cleaning the ion sources • Replacing filaments • Replacing the EM horn Agilent 7000/7010 Series Triple Quad GC/MS Troubleshooting and Maintenance Manual provides information about symptoms that indicate the need for analyzer maintenance. The troubleshooting material in the online help in the MassHunter Workstation software provides more extensive information.
  • Page 129 Do not touch the HED ceramic insulator. More information is available If you need more information about the locations or functions of analyzer components, refer to the Agilent 7000/7010 Series Triple Quad GC/MS Troubleshooting and Maintenance Manual. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 130: To Open The Front Analyzer Chamber

    Do not touch any part until you are sure it is cool. Always wear clean gloves to prevent contamination when working in the analyzer C A U T I O N chamber. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 131 C A U T I O N Verify that the MS is vented. Verify that both the front and rear side plate screws are completely loose. Gently swing the side plate out. (See Figure 40 on page 132.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 132 General Maintenance Quad driver boards Front analyzer side plate Chamber closed Thumbscrews Post-filter Side plate Ceramic source board Ion source Front analyzer Figure 40 The front analyzer chamber on a 7010 MS 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 133: To Remove The Ei Hes

    When disconnecting leads, pull on the connectors, not on the wires. C A U T I O N Open the front analyzer chamber. (See “To Open the Front Analyzer Chamber” on page 130.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 134 The source contacts have spring loaded pins so some force must be applied to pull the source out. (See Figure 41.) Ion source Thumbscrew Ceramic source board Source finger grip Thumbscrew Source heater and temperature sensor wires Source radiator Figure 41 The EI HES before removal 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 135: To Connect Or Disconnect Wiring To The Ei Hes

    Use tweezers or pliers to connect or disconnect the ceramic board wire leads (red, white, black, and gray) at the source connectors. (See Figure 42.) Ceramic source board Repeller Filament common Filament 1 Filament 2 Figure 42 Wiring for the EI HES 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 136: To Remove The Ei Xtr Source

    When disconnecting leads, pull on the connectors, not on the wires. C A U T I O N Open the front analyzer chamber. (See “To Open the Front Analyzer Chamber” on page 130.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 137 Remove the thumbscrews that hold the ion source in place. (See Figure 43.) Pull the ion source out of the source radiator. Ceramic source board Ion source Thumbscrews Source heater and temperature sensor wires Source radiator Figure 43 Removing the EI XTR source 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 138: To Connect Or Disconnect Wiring To The Ei Xtr Source

    Use tweezers or pliers to connect or disconnect the source heater wire leads at the CSB. (See Figure 44.) Ceramic source board Entrance lens Ion focus lens Filament 1 Repeller Filament 2 Extractor lens Source heater Source RTD Figure 44 Wiring for the EI XTR 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 139: To Disassemble The Ei Hes

    Use a T6 Torx screwdriver to remove the screw and locking ring for the lens insulator that secures the lens stack in the source body, then remove the lens stack. 10 If necessary, use gravity to remove the lens stack ceramic insulator from the source body. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 140 11 Remove the five lenses from the lens insulator/holder. Figure 45 Exploded parts view of the EI HES Table 14 Parts list for EI HES (Figure Item number Item description Part number Source finger grip G7002-20008 Filament block G7002-20019 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 141 High efficiency dual filament G7002-60001 Ring heater/sensor assembly G7002-60043 Source mount 1.5 mm G7002-60053 Repeller assembly G7002-60057 Not shown HES assembly G7004-67055 * The number in parenthesis is the number engraved on the lens 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 142: To Disassemble The Ei Xtr Source

    Remove the setscrew securing the lenses to the source body. Pull the lenses out of the source body and separate the lens insulator, ion focus lens, extractor lens, extractor lens insulator, and extended entrance lens. (See Figure 46 on page 143.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 143 General Maintenance Figure 46 Disassembling the EI XTR source Table 15 Parts list for EI XTR source (Figure Item Description Part number Setscrews G3870-20446 Screws G3870-20021 Source body G3870-20440 Extractor lens G3870-20444 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 144 Ion focus lens 05971-20143 Repeller insulator G1099-20113 Repeller G3870-60171 Flat washer 3050-0891 Belleville spring washer 3050-1301 Repeller nut 0535-0071 Source heater block assembly G3870-60177 Repeller block insert G3870-20135 Not shown EI XTR source assembly G7003-67720 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 145: To Clean The Ei Hes

    Post extractor lens 2 (3) • Ion focus lens (2) • Extended entrance lens (1) These are the parts that contact the sample or ion beam. The other parts normally should not require cleaning. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 146 If the contamination is serious, such as an oil backflow into the analyzer, seriously consider replacing the contaminated parts. Do not use the abrasive slurry on the source mount bushings. C A U T I O N 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 147 WA R N I N G precautions. Place the parts in a clean beaker. Loosely cover the beaker with clean aluminum foil (dull side down). Dry the cleaned parts in an oven at 100 °C for 5–6 minutes. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 148: To Clean The Ei Xtr Source

    Source body • Repeller block insert • Extractor lens • Ion focus lens • Entrance lens These are the parts that contact the sample or ion beam. The other parts normally should not require cleaning. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 149 Use enough force to remove all discolorations. Polishing the parts is not necessary; small scratches will not harm performance. Also abrasively clean the discolorations where electrons from the filaments enter the source body. 3 Rinse away all abrasive residue with reagent-grade methanol. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 150 Take care to avoid contaminating cleaned and dried parts. Put on new, clean gloves before N O T E handling the parts. Do not set the cleaned parts on a dirty surface. Set them only on clean, lint-free cloths. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 151: To Assemble The Ei Hes

     It is easier to insert lens 5 on an angle because the lens stack at this point makes the lens insulator less flexible. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 152 Using the T6 Torx driver, install and secure the gold plated setscrew and lock ring lens insulator that holds the lens assembly in place. (See Figure 50.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 153 Fully insert the three filament leads into the filament block. 11 Inset the filament block into the source holder, and use a T6 Torx screwdriver to secure it to the holder with the gold plated screw. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 154 Source finger grip G7002-20008 Filament block G7002-20019 Extractor lens (5)*, with 3 mm opening G7002-20061 Ceramic insulator for extractor G7002-20064 Entrance lens assembly, Extended, HES (1)* G7002-20165 Ion focus lens (2)* G7002-20068 Lens insulator/holder G7002-20074 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 155 Use care when inserting the lenses into the lens insulator casing. Putting excessive C A U T I O N stress on this casing can break or crack it. If this happens, do not attempt to operate with a defective lens insulator, it must be replaced. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 156: To Assemble The Ei Xtr Source

    Attach the repeller assembly to the source body with two screws and washers. Install the filaments. (See “To Install a Filament on the EI XTR Source” page 165.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 157 Assembling the EI XTR source Table 17 Parts list for EI XTR source (Figure Item Description Part number Setscrews G3870-20446 Screws G3870-20021 Source body G3870-20440 Extractor lens G3870-20444 Extractor lens insulator G3870-20445 Filaments G7005-60061 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 158 Ion focus lens 05971-20143 Repeller insulator G1099-20113 Repeller G3870-60171 Flat washer 3050-0891 Belleville spring washer 3050-1301 Repeller nut 0535-0071 Source heater block assembly G3870-60177 Repeller block insert G3870-20135 Not shown EI XTR source assembly G7003-67720 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 159: To Install The Ei Hes

    (See Figure 53). Slide the ion source into the source radiator until resistance is felt. EI HES thumbscrew attachment Slots for source mount alignment EI HES thumbscrew attachment Figure 53 The EI HES radiator 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 160 “To Close the Front Analyzer Chamber” on page 169.) Pumpdown the MS. (See “To Pumpdown the MS” on page 95.) Tune the MS. (See “To Autotune the MS for EI Mode” on page 92.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 161: To Install The Ei Xtr Source

    Install and hand-tighten the source thumbscrews. Do not overtighten the thumbscrews. (See Figure 54 on page 162.) Connect the ion source wires. (See “To Connect or Disconnect Wiring to the EI XTR Source” on page 138.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 162 “To Close the Front Analyzer Chamber” on page 169.) Pumpdown the MS. (See “To Pumpdown the MS” on page 95.) Tune the MS. (See “To Autotune the MS for EI Mode” on page 92.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 163: To Remove A Filament From The Ei Xtr Source

    The analyzer, GC/MS interface, and other components in the analyzer chamber WA R N I N G operate at very high temperatures. Do not touch any part until you are sure it is cool. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 164 General Maintenance Ion source body Screw Filament Figure 55 Changing the EI XTR filament 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 165: To Install A Filament On The Ei Xtr Source

    “To Close the Front Analyzer Chamber” on page 169.) Pumpdown the MS. (See “To Pumpdown the MS” on page 95.) Autotune the MS. (See “To Autotune the MS for EI Mode” on page 92.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 166: To Remove The Ei Hes Filaments

    Open the front analyzer chamber. (See “To Open the Front Analyzer Chamber” on page 130.) Remove the ion source, and place it on a clean cloth on your work surface. (See “To Remove the EI HES” on page 133.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 167 Remove the dual filament from the filament block by lifting the source body up off of the filament block, while holding the filament block so that the dual filament will not fall and become damaged. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 168: To Install The Ei Hes Filaments

    “To Close the Front Analyzer Chamber” on page 169.) Pumpdown the MS. (See “To Pumpdown the MS” on page 95.) Autotune the MS. (See “To Autotune the MS for EI Mode” on page 92.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 169: To Close The Front Analyzer Chamber

    C A U T I O N pumpdown. Do not use a screwdriver to tighten the thumbscrew. Once the MS has pumped down, close the left analyzer cover and replace the window cover. Tune the MS. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 170: To Remove Left Rear Cover

    Lift the bottom flap of the cover out of the groove in the back of the MS to free the cover. Do not remove any other covers. Dangerous voltages are present under the other WA R N I N G covers. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 171 General Maintenance Analyzer window cover Analyzer cover Rear cover Figure 57 Removing covers 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 172: To Open The Rear Analyzer Chamber

    Always wear clean gloves to prevent contamination when working in the analyzer C A U T I O N chamber. Loosen the rear analyzer side plate thumbscrews, if they are fastened. (See Figure 40 on page 132.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 173 In the next step, if you feel resistance, stop. Do not try to force the side plate open. C A U T I O N Verify that the MS is vented. Verify that both the front and rear side plate screws are completely loose. Gently swing the side plate out. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 174 General Maintenance Rear analyzer side plate Detector board Thumbscrews Chamber closed Collision cell Rear analyzer Detector Rear side plate Chamber open Figure 58 The rear analyzer chamber 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 175: To Replace The Em Horn

    Hold the new horn with blue signal wire end down, and attach the signal wire to the connector in the sideplate. (See Figure 60 on page 176.) Slide the EM horn into position. Close the retaining clip. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 176 “To Close the Rear Analyzer Chamber” on page 177.) EM horn Retaining clip Blue signal wire Figure 59 Replacing the EM horn (on a series II detector) Figure 60 EM horn for a series II detector 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 177: To Close The Rear Analyzer Chamber

    Do not use a screwdriver to tighten the thumbscrew. If hydrogen or any other flammable material is used for the carrier gas, gently hand-tighten the top thumbscrew on the rear analyzer side plate. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 178 General Maintenance Once the MS has pumped down, close the left analyzer cover, replace the rear cover, and replace the analyzer window cover. Tune the MS. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 179 To Install the CI/EI XTR Source Radiator To Switch from the CI Source to the EI HES To Install a CI Source Filament To Remove the CI Source Radiator from the 7010 Series TQ To Disassemble the CI Source To Clean the CI Source...
  • Page 180: Ci Maintenance

    • Ensure the reagent gas inlet line(s) are equipped with gas purifiers (not applicable for ammonia). • Use extra-high purity reagent gases; 99.99% or better for methane and as pure as is available for other reagent gases. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 181: To Switch From An Ei Xtr Source To The Ci Source

    (See“Electrostatic discharge” on page 126.) Take antistatic precautions before you open the analyzer chamber. Install the CI source. (See “To Install the 7010 Series CI Source” page 185.) Close the analyzer. (See “To Close the Front Analyzer Chamber” page 169.) Pumpdown and switch from EI to CI mode.
  • Page 182: To Remove The Ei Hes Radiator

    “To Connect or Disconnect EI HES Radiator Wiring” on page 184.) Use a T10 Torx screwdriver to loosen the two screws that secure the radiator to the analyzer and place the radiator in its storage container. (See Figure 61.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 183 CI Maintenance Radiator screw HES radiator Radiator screw Figure 61 Removing the EI HES radiator 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 184: To Connect Or Disconnect Ei Hes Radiator Wiring

    RTD wires from the CSB. Ceramic source board Entrance lens Ion focus lens Post extractor lens 2 Post extractor lens Extractor lens Ground wire Heater wires RTD wires Figure 62 HES radiator wiring 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 185: To Install The 7010 Series Ci Source

    CI Maintenance To Install the 7010 Series CI Source Electrostatic discharges to analyzer components are conducted to the side board C A U T I O N where they can damage sensitive components. Wear a grounded antistatic wrist strap, ...
  • Page 186: To Install The 7000 Series Ci Source

    Installing the CI source Close the analyzer door. (See “To Close the Front Analyzer Chamber” page 169.) Pumpdown the MS. (See “To Pumpdown the MS” on page 95.) Tune the MS. (See “CI Autotune” on page 109.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 187: To Switch From The Ci Source To The Ei Hes

    Remove the CI source radiator. The Radiator should not be removed or installed with the CI source in place. (See “To Remove the CI Source Radiator from the 7010 Series TQ” on page 201.) Place the CI source radiator in the storage container.
  • Page 188: To Install The Ei Hes Radiator

    Figure 65 on page 189.) Attach the wires to the radiator or CSB. Do not bend the wires any more than necessary. (See “To Connect or Disconnect EI HES Radiator Wiring” on page 184.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 189 CI Maintenance Radiator screw HES radiator Radiator screw Figure 65 Securing the EI HES radiator 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 190: To Switch From The Ci Source To The Ei Xtr Source

    Wear a grounded antistatic wrist strap, and take other antistatic precautions before you open the analyzer chamber.  (See “Electrostatic discharge” on page 126.) Pumpdown the MS. (See “To Pumpdown the MS” on page 95.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 191: To Connect Or Disconnect 7010 Series Ci Source Wiring

    CI Maintenance To Connect or Disconnect 7010 Series CI Source Wiring This topic is used to identify the wiring connections to the ion source. (For detailed procedures and videos, see “To Remove the 7000 Series CI Source” page 194 or “To Install the 7010 Series CI Source”...
  • Page 192: To Connect Or Disconnect 7000 Series Ci Source Wiring

    Use tweezers or pliers to connect or disconnect the source heater wire leads at the CSB. Ceramic source board Entrance lens Ion focus Filament 1 Repeller Dummy filament Source heater Source RTD Figure 67 Wiring between the CSB and the CI source 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 193: To Switch From The Ci Source To An Ei Xtr Source

    If installing an EI SS or Inert source, remove the interface tip seal and put it in the CI storage container. Pumpdown the MS. (See “To Pumpdown the MS” on page 95.) Load your EI tune file and Tune the MS. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 194: To Remove The 7000 Series Ci Source

    7000 Series CI Source Wiring” on page 192.) Remove the two large thumbscrews that hold the ion source in place. Pull the ion source out of the source radiator, and place it in its storage container. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 195: To Remove The 7010 Series Ci Source

    CI Maintenance To Remove the 7010 Series CI Source Materials needed • Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029) • Tweezers (8710-2460)   Procedure Vent the MS. (See “To Vent the MS” on page 99.) The analyzers, GC/MS interface, and other components in the analyzer chamber WA R N I N G operate at very high temperatures.
  • Page 196: To Switch From The Ei Hes To The Ci Source

    1 to 2 mm. (See “To Install the 7010 Series CI Source” on page 185.) 12 Pumpdown and switch from EI to CI mode. (See “To Pumpdown and Switch from EI to CI Mode” on page 118.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 197: To Install The Ci/Ei Xtr Source Radiator

    When disconnecting leads, pull on the connectors, not on the wires. C A U T I O N Align the radiator over the two guide pins on the analyzer support, and secure it using the two retained screws with a number T10 Torx screwdriver. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 198 CI Maintenance Connect the green ground wire to the radiator. (See Figure 68.) Radiator Green ground wire Figure 68 CI Source radiator 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 199: To Switch From The Ci Source To The Ei Hes

    “To Open the Front Analyzer Chamber” on page 130.) Remove the CI source. (See “To Remove the 7010 Series CI Source” on page 195.) Place the CI source in the storage container. Remove the CI source radiator. The Radiator should not be removed or installed with the CI source in place.
  • Page 200: To Install A Ci Source Filament

    Secure the filament to the ion source body with the screw. After installing the filament, verify that it is not grounded to source body. Reinstall the CI source. (See “To Install the 7010 Series CI Source” page 185 or see “To Remove the CI Source Filament”...
  • Page 201: To Remove The Ci Source Radiator From The 7010 Series Tq

    CI Maintenance To Remove the CI Source Radiator from the 7010 Series TQ Materials needed • Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029) • Tweezers (8710-2460)   • T10 Torx screwdriver (5182-3466) Procedure The analyzers, GC/MS interface, and other components in the analyzer chamber WA R N I N G operate at very high temperatures.
  • Page 202 Use a T-10 Torx screwdriver to loosen the two retained screws that secure the radiator to the analyzer and place the radiator in its storage container. Radiator Green ground wire Figure 69 CI/EI XTR Source radiator 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 203: To Disassemble The Ci Source

    Remove the setscrew securing the lenses to the source body. Pull the lenses out of the source body and separate the lens insulator, ion focus lens, drawout cylinder, drawout lens, and entrance lens. (See Figure 70 on page 204.) 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 204 Table 18 Parts list for CI source (Figure Item Description Part number Set screw G1999-20022 Filament screw G1999-20021 CI repeller insulator G1999-20433 CI lens insulator G3170-20540 CI drawout cylinder G1999-20444 CI drawout plate G1999-20446 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 205 Washer spring curved 2.2 mm-ID 4.5 mm-OD, Qty. 2 3050-1374 Flat washer 3050-9082 Not shown Package, GC/MS source Clamshell G7002-80008 Not shown Bracket, GC/MS source, Clamshell G7002-00008 Not shown CI source assembly G7002-67404 Not shown CI source assembly (without tip seal) G7077-67404 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 206: To Clean The Ci Source

    Source body • Drawout plate • Drawout cylinder • Ion focus lens • Entrance lens These are the parts that contact the sample or ion beam. The other parts normally should not require cleaning. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 207 Rinse away all abrasive residue with reagent-grade methanol. Make sure all abrasive residue is rinsed before ultrasonic cleaning. If the methanol becomes cloudy or contains visible particles, rinse again three times. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 208 Take care to avoid contaminating cleaned and dried parts. Put on new, clean gloves before N O T E handling the parts. Do not set the cleaned parts on a dirty surface. Set them only on clean, lint-free cloths. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 209: To Assemble The Ci Source

    The nut should only be finger-tight. Place the heater block assembly on top of the source body. Reinstall the dummy filament, and the filament and attach with the setscrews. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 210 Table 19 Parts list for CI source (Figure Item Description Part number Set screw G1999-20022 Filament screw G1999-20021 CI repeller insulator G1999-20433 CI lens insulator G3170-20540 CI drawout cylinder G1999-20444 CI drawout plate G1999-20446 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 211 Washer spring curved 2.2 mm-ID 4.5 mm-OD, Qty. 2 3050-1374 Flat washer 3050-9082 Not shown Package, GC/MS source Clamshell G7002-80008 Not shown Bracket, GC/MS source, Clamshell G7002-00008 Not shown CI source assembly G7002-67404 Not shown CI source assembly (without tip seal) G7077-67404 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 212: To Remove The Ci Source Filament

    The analyzer, GC/MS interface, and other components in the analyzer chamber WA R N I N G operate at very high temperatures. Do not touch any part until you are sure it is cool. 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 213 CI Maintenance CI source body Filament Screw Figure 73 Changing the CI source filament 7000/7010 Series TQ GC/MS Operating Manual...
  • Page 214: To Install A Ci Source Filament

    Secure the filament to the ion source body with the screw. After installing the filament, verify that it is not grounded to source body. Reinstall the CI source. (See “To Install the 7010 Series CI Source” page 185.) Pumpdown the MS. (See “To Pumpdown the MS”...
  • Page 216 Agilent Technologies ‘ © Agilent Technologies, Inc. Printed in USA, July 2017 *G7003-90046* *G7003-90044* G7003-90046...