Figure 15 Default Setup Screen - Agilent Technologies InfiniiVision 7000 Series Service Manual

Hide thumbs Also See for InfiniiVision 7000 Series:
Table of Contents

Advertisement

4
Troubleshooting
82
Vertical - Channel 1 on, 5 V/div scale, dc coupling, 0 V
position, probe factor to 1.0 if an AutoProbe probe is
not connected to the channel
Trigger - Edge trigger, Auto sweep mode, 0 V level,
channel 1 source, dc coupling, rising edge slope, 60 ns
holdoff time
Display - Vectors on, 33% grid intensity, infinite
persistence off
Other - Acquire mode normal, Run/Stop to Run, cursor
measurements off
Default setup screen
Figure 15
3 If your screen looks substantially different, replace the
system board.
7000 Series Oscilloscopes Service Guide