Figure 3. Setting Up Equipment For Voltage Measurement Accuracy - Agilent Technologies InfiniiVision 7000 Series Service Manual

Hide thumbs Also See for InfiniiVision 7000 Series:
Table of Contents

Advertisement

2
Testing Performance
36
Oscilloscope
Oscilloscope
Calibrator
BNC Tee
BNC (f) to dual
banana adapter
Setting up Equipment for Voltage Measurement Accuracy
Figure 3
Test
4 Adjust the output so that the multimeter reading displays
the first Volts/div calibrator setting value in
Wait a few seconds for the measurement to settle.
5 Press the Y2 softkey, then position the Y2 cursor to the
center of the voltage trace using the Entry knob.
The Y value on the lower line of the display should be
within the test limits of
within the test limits, go to the "Troubleshooting"
chapter. Then return here.
6 Continue to check the voltage measurement accuracy with
the remaining Volts/div setting values in
Digital
Multimeter
Table
7. If a result is not
Table
7000 Series Oscilloscopes Service Guide
Table
7.
7.