2
Testing Performance
Horizontal t Accuracy
Generator Setting
Period
100 s
Period
100 ns
Period
5 ns
Trigger Sensitivity
Internal trigger
1 GHz models
500 MHz models
350 MHz models
100 MHz models
Trigger Sensitivity
External trigger (7052, 7032, 7012)
± 1 V range:
7052
7032
7012
External trigger (7052, 7032, 7012)
± 8 V range:
7052
7032
7012
External trigger (7104, 7054, 7034, 7014)
* Where applicable
58
Test Limits
99.8 s to 100.2 s
99.8 ns to 100.2 ns
4.96 ns to 5.04 ns
Test Limits
1 division at 1 GHz
1 division at 500 MHz
1 division at 350 MHz
1 division at 100 MHz
Test Limits
100 mV at 25 MHz
200 mV at 500 MHz
200 mV at 350 MHz
200 mV at 100 MHz
250 mV at 25 MHz
500 mV at 500 MHz
500 mV at 350 MHz
500 mV at 100 MHz
500 mV at 500 MHz
Results
________
________
________
Channel 1 Channel 2 Channel 3* Channel 4*
________ ________ ________ _______
________ ________ ________ _______
________ ________ ________ _______
________ ________ ________ _______
Channel 1 Channel 2 Channel 3* Channel 4*
________ ________ ________ ________
________ ________ ________ ________
________ ________ ________ ________
________ ________ ________ ________
________ ________ ________ ________
________ ________ ________ ________
________ ________ ________ ________
________ ________ ________ ________
________ ________ ________ ________
7000 Series Oscilloscopes Service Guide

Advertisement

Table of Contents
loading

Table of Contents