3 In the Waveform Generator Settings Menu, press the Modulation softkey.
4 In the Waveform Generator Modulation Menu:

To set up Amplitude Modulation (AM)

In the Waveform Generator Modulation Menu (under [Wave Gen] > Settings >
Modulation):
1 Press the Type softkey and turn the Entry knob to select Amplitude
2 Press the Waveform softkey and turn the Entry knob to select the shape
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
Press the Modulation softkey to enable or disable modulated waveform
generator output.
You can enable modulation for all waveform generator function types
except pulse, DC, and noise.
Press the Type softkey and turn the Entry knob to select the
modulation type:
Amplitude Modulation (AM) — the amplitude of the original carrier
signal is modified according to the amplitude of the modulating
signal. See
"To set up Amplitude Modulation (AM)"
Frequency Modulation (FM) — the frequency of the original carrier
signal is modified according to the amplitude of the modulating
signal. See
"To set up Frequency Modulation (FM)"
Frequency-Shift Keying Modulation (FSK) — the output frequency "shifts"
between the original carrier frequency and a "hop frequency" at
the specified FSK rate. The FSK rate specifies a digital square
wave modulating signal. See
Modulation (FSK)"
Modulation (AM).
of the modulating signal:
Sine
Square
Ramp
Sine Cardinal
"To set up Frequency- Shift Keying
on page 214.
Waveform Generator
on page 211.
on page 213.