Delay

Delay measures the time difference from the selected edge on source 1
and the selected edge on source 2 closest to the trigger reference point at
the middle threshold points on the waveforms. Negative delay values
indicate that the selected edge of source 1 occurred after the selected
edge of source 2.
Source 1
Source 2
1 Press the [Meas] key to display the Measurement Menu.
2 Press the Source softkey; then turn the Entry knob to select the first
3 Press the Type: softkey; then, turn the Entry knob to select Delay.
4 Press the Settings softkey to select the second analog channel source
5 Press the
6 Press the Add Measurement softkey to make the measurement.
The example below shows a delay measurement between the rising edge of
channel 1 and the rising edge of channel 2.
184
analog channel source.
and slope for the delay measurement.
The default Delay settings measure from the rising edge of channel 1 to
the rising edge of channel 2.
Back/Up key to return to the Measurement Menu.
Delay
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide