Preshoot - Agilent Technologies InfiniiVision 2000 X-Series User Manual

Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

Figure 34

Preshoot

Preshoot is distortion that precedes a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being
measured (edge closest to the trigger reference point).
Rising edge preshοot =
Falling edge preshοot = Base − D local Minimum
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
Automatic Overshoot measurement
local Maximum − D Top
Amplitude
Amplitude
Measurements
× 100
× 100
13