Connect Probes To The Oscilloscope; Maximum Input Voltage At Analog Inputs; Do Not Float The Oscilloscope Chassis - Agilent Technologies InfiniiVision 2000 X-Series User Manual

Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

1
Getting Started
2 The oscilloscope automatically adjusts for input line voltages in the
Always use a grounded power cord. Do not defeat the power cord ground.
WA R N I N G
3 Press the power switch.
The power switch is located on the lower left corner of the front panel.
The oscilloscope will perform a self- test and will be operational in a few
seconds.

Connect Probes to the Oscilloscope

1 Connect the oscilloscope probe to an oscilloscope channel BNC
2 Connect the probe's retractable hook tip to the point of interest on the
C A U T I O N
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
With N2862A or N2863A 10:1 probe: 300 Vrms
C A U T I O N
Defeating the ground connection and "floating" the oscilloscope chassis will probably
result in inaccurate measurements and may also cause equipment damage. The probe
ground lead is connected to the oscilloscope chassis and the ground wire in the power
cord. If you need to measure between two live points, use a differential probe with
sufficient dynamic range.
24
range 100 to 240 VAC. The line cord provided is matched to the country
of origin.
connector.
circuit or device under test. Be sure to connect the probe ground lead
to a ground point on the circuit.

Maximum input voltage at analog inputs

Do not float the oscilloscope chassis

Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide