Setting the Head Installation

Set how the Sensor Head is installed.
FUN Mode-[SENSING]-[SETTING]
Setting
DIFFUSE
Select this item when the Sensor Head is installed for diffuse reflection sensing.
REGULAR
Select this item when the Sensor Head is installed for regular reflection sensing.

Setting the Emitted Light Intensity

Set the amount of light emitted from the Sensor Head to match the state of the workpiece
surface.
The response may slow down if workpieces having differing reflection factors such as black-and-white
workpieces are measured with the FUN mode set to [AUTO]. In this case, narrow the adjustment range
by setting [RANGE]. If this does not increase the response speed to keep up with measurement, select
[FIXED].
FUN Mode-[SENSING]-[LASER]
Setting
AUTO
RANGE
UPPER
LOWER
FIXED
Description
ZS-LD20T/20ST/40T
ZS-LD20T/20ST/40T
Automatically adjusts the emitted light amount according to the reflection
factor of the workpiece. Note that the response time varies with each mea-
surement. (default value)
Sets the adjustment range for the AUTO setting.
If the response speed cannot keep up with measurement at the default
value, narrow the range.
Range: 0.1 to 80% (default: LOWER 0.1%, UPPER 80%)
Level
0.1%
·
·
80%
Sets the emitted light amount to a fixed value.
For reference levels, see [RANGE].
Range: 0.1% to 80%.
Setting the Sensing Conditions
ZS-LD50/50S/80/130/200/350S
ZS-LD50/50S/80/130/200
Description
Sensitivity
Color of workpiece
Low
High
Section 3
Bright
Dark
ZS-L
3-13
User's Manual