Q&A
Question
When scaling is executed, an error appears and
settings cannot be made.
When focusing is executed, an error appears and
settings cannot be made.
Is warm-up operation still required even if LD-
OFF input is cancelled?
Answer
Scaling cannot be set for one of the following reasons:
• Scaling has been attempted when the measured value is
outside the measuring range.
• When two-point scaling has been executed, the distance
between the measured values for the two points is not 1%
or more of the rated measuring range.
p.3-20
Monitor focus settings cannot be made when the distance
between the two specified points is not 1% or more of the
rated measuring range.
p.4-2
Yes. As when the power is turned ON, warm-up operation
must be performed for about 30 minutes.
Section 6
APPENDIX
ZS-L
6-5
User's Manual