MAC 5500 ST Requirements and Configuration

Compatible Blood Pressure Units

Revision E
Installation: MAC 5500 ST Requirements and Configuration
Following is a list of interface requirements and setup configurations required for
the devices listed when used with the MAC 5500 ST option.
Colin - Model ST-780
Connection Requirements - Use cable PN 2008112-001 to connect from the
MAC5500 port 1 to the Colin serial port. Use cable PN 2008111-001 to connect
from the MAC 5500 ANA/TTL port to the Colin QRS trigger input.
Device Configuration Requirements - None
MAC 5500 Configuration Requirements – At the Main Menu complete the
following in the order shown below:
Select System Setup,
Enter System password,
Exercise Test,
Inputs/Outputs,
Change Blood Pressure to Nipon-Colin.
Sun Tech - Model Tango
Connection Requirements - Use cable PN 2008113-001 to connect from the MAC
5500 port 1 to the Sun Tech serial port. Use cable PN 2008111-001 to connect from
the MAC5500 ANA/TTL port to the Sun Tech QRS trigger input.
Device Configuration Requirements – At the Tango Main Menu complete the
following in the order shown below:
Select Utilities,
Select Device,
Scroll to ECG Trigger and press enter,
Scroll to DIGITAL
Scroll to EXIT and press Enter,
Scroll to Test Parameters and press Enter,
With Technique highlighted, press Enter,
Scroll to DKA and press Enter,
Scroll to EXIT and press Enter,
Scroll to EXIT and press Enter to return to the display screen.
MAC 5500 resting ECG analysis system
2020299-020
and press enter,
3-15