Item
Frequency Response/Input
Frequency Response/Output
Upper Limit
Lower Limit
Sensitivities
Raw Data and Template
Average Beat
Filtered Signals and Vector
Magnitude
Analysis Sampling Rate
Digital Sampling Rate
High/Low Pass Filters
ADC Resolution
Analysis Resolution
Item
Height
Width
Depth
Weight
1. without trolley
Item
Operating Conditions
Temperature
Relative Humidity
Atmosphere Pressure
Revision E
Appendix B – Technical Specifications: Technical Specifications

Hi-Res and PHi-Res Signal-Averaged Electrocardiography

-3 dB @ 0.01 and 250 Hz
250 Hz
0.01, 25, 40 or 80 Hz
20 mm/mV
20 mm/mV and 50 mm/mV
1 mm/µV
1000 samples per second per channel
4000 samples per second per channel
Spectral filter using Fast Fourier Transform (FFT)
1.22 µV/LSB
0.1525 µV/LSB

Physical

9.4 cm (3.7 in) with display closed
38.1 cm (15.0 in)
35.1 cm (13.8 in)
6.8 kg (15 lb) without paper

Environmental

10° C to 40° C (50° F to 104° F)
20% to 95% RH noncondensing
700 to 1060 hPa