Type-S Trolley Disassembly; Power Supply; Removal - GE MAC 5500 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 5500:
Table of Contents

Advertisement

Type-S Trolley Disassembly

Power Supply

Removal

5-8
Maintenance: Disassembly Guidelines
To dismount the MAC 5500 from the Type-S trolley, follow the steps shown in the
following illustration.
NOTE
A #10 TORX driver is required for disassembly and assembly of the power
supply.
1.
Turn the unit over so the bottom side is up.
2.
Using a #10 TORX driver, remove the three screws holding the power supply in
place.
3.
Lift the power supply to expose the wiring harness and ground wire.
4.
Remove P2 from J2 on the power supply assembly and the ground wire
connection from the power supply chassis.
MAC 5500 resting ECG analysis system
2020299-020
39A
Revision E