3-16
Installation: MAC 5500 ST Requirements and Configuration
MAC 5500 Configuration Requirements – At the Main Menu complete the
following in the order shown below:
Select System Setup,
Enter System password,
Exercise Test,
Inputs/Outputs,
Change Blood Pressure to Suntech.
Ergoline - Model Ergoline 900
Connection Requirements – Use cable PN 2008110-001 to connect from the MAC
5500 port 1 to the Ergoline serial port. Use cable PN 2008115-001 to connect from
the MAC5500 ANA/TTL port to the Ergoline QRS trigger input.
Device Configuration Requirements – See Ergoline 900 Operator's Manual.
MAC 5500 Configuration Requirements – At the Main Menu complete the
following in the order shown below:
Select System Setup,
Enter System password,
Exercise Test,
Inputs/Outputs,
Change Blood Pressure to Ergoline Ergometer.
Model T2000
Connection Requirements – Use cable PN 2007918-001 (T2000) to connect from
the MAC 5500 port 1 to the treadmill serial port.
Device Configuration Requirements – None.
MAC 5500 Configuration Requirements – Use the Edit Protocol application to set
the protocol Test Type to Treadmill in MPH or Treadmill in Km/H for protocols that
will be used with this treadmill.
Connection Requirements – There are no cables available from GE Medical
Systems Information Technologies to interface to analog treadmills. The customer is
responsible for making the appropriate cable. Speed and grade signals for
controlling analog treadmills are available on pins 2 (Slow Analog Output) and 8
(Fast Analog Output) of the ANA/TTL port. Pins 1, 4 and 5 are tied to ground.
Device Configuration Requirements – None.
MAC5500 Configuration Requirements – Use the Edit Protocol application to set
the protocol Test Type to Analog Treadmill in MPH or Analog Treadmill in Km/H
for protocols that will be used with this treadmill.
Configure pin 2 on the ANA/TTL port by selecting the following:
MAC 5500 resting ECG analysis system
2020299-020
Revision E