Revision E
Maintenance: Disassembly Guidelines
5.
Enter the IP Address, Subnet Mask, and Gateway provided by the network
administrator
6.
Enter the Port Number used by the MUSE system.
7.
Select Return and press Enter.
The Network Settings window closes.
8.
On the Basic System menu, select Transmission and press Enter.
The Transmission window opens.
9.
Enter the following information.
Table 2. Transmission Information
Field
Use IR for serial line
Serial line baud rate
Default location
10. Select Return and press Enter.
The Transmission window closes.
11. On the Basic System menu, select Return and press Enter.
The Basic System menu closes.
12. On the System Setup menu, select Save Setup >To System.
This saves the settings to the communication board. This should take
approximately 8 to 10 seconds. If it takes less than 8 seconds, the settings were
not saved successfully. Repeat a second time.
13. Cycle power on the MAC 3500/MAC 5500 system.
This ensures the network settings are reread from the communications board.
14. Wait at least 10 minutes and verify the MAC 3500/MAC 5500 can
communicate with the MUSE system.
Windows File Server 2000/2003 has an ARP cache that regenerates the MAC
Address/Physical Address table every 10 minutes. When you replace the
communications board, you change the MAC Address/Physical Address. By
waiting at least 10 minutes, you allow the MUSE operating system to
regenerate the MAC Address/Physical Address cache so the MAC 3500/MAC
5500 is able to communicate again with the MUSE system.
15. Perform all necessary functional checkouts.
MAC 5500 resting ECG analysis system
2020299-020
Value
No
115.2K
Ethernet (MUSE)
5-21