Service Only Setups - GE MAC 5500 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 5500:
Table of Contents

Advertisement

Service Only Setups

5-16
Maintenance: Disassembly Guidelines
Programming Primary Boot...
Programming Over
System is Shutting Down
If the boot code is being downgraded to an earlier version, a message
similar to the following message displays.
Press 'Enter' to start Installation
Press Enter. The following messages display.
Programming Primary Boot...
Programming Over
System is Shutting Down
8.
Reboot the system and verify the software version.
1.
From the Main Menu, select System Setup.
2.
Press Shift + F2 at the System Setup menu.
3.
Enter the service password and press Enter.
4.
The Service Only Setup window is displayed.
Service Only Setup
Serial number:
Update Primary Boot:
Print head resistance:
Keyboard:
Return
5.
Enter the serial number of the system.
NOTE
This is the number which was used when the option codes for this system
were generated. The number entered here must match the serial number on
MAC 5500 resting ECG analysis system
2020299-020
Current Boot Version:
New Boot Version:
_ _ _ _ _ _ _ _ _ _ _ _ _
_ _ _
_ _ _
_ _ _ _ _ _ _ _ _
Revision E