Display/Keyboard Reassembly

Reassembly

5-12
Maintenance: Disassembly Guidelines
Display Ground
LCD Flex Cable
Backlight
Flex Cable
Flex Cable
Slots
1.
Insert both flex cables through flex cable slots and position them as shown.
2.
Tilt the display/keyboard assembly to the left and with the roll pin of the hinge
(metal rod) parallel to the left hinge base, insert the rod into the left hinge base
and lower the display/keyboard assembly in place.
3.
Slide tabs into their mounting slots and set the display/keyboard assembly in
place.
4.
Connect the three cables from the display/keyboard assembly to the main PCB.
Be sure to lift the locks up prior to attempting to insert the cables into the
connectors.
5.
Slide the display hinge (metal rod) to the right until it locks into the right hinge
base.
6.
Replace the hinge bracket with the two TORX screws removed earlier.
7.
Replace the screw and display ground at the left of the hinge rod.
8.
Replace the two TORX mounting screws on the right side of assembly.
1.
Raise the display to the vertical position.
2.
Make sure the bezel surrounding the rear panel connectors is in place. Make
sure the release mechanism for the Smartmedia card functions properly.
3.
Lower the top cover down around the display and set in position.
4.
Snap the rear of the top cover in place and then, gently pulling on the thin
plastic strip at the front of the top cover, position it in place under the keyboard
assembly.
5.
Replace the screws removed in disassembly.
MAC 5500 resting ECG analysis system
Fasten to writer as shown
LCD Flex Cable
Backlight Flex Cable
Align Flex to Pin
45A
2020299-020
Display Ground
Roll Pin
Revision E
44A