Battery and Patient Cable Replacement
Battery Replacement
Patient Cable Replacement
5-6
Maintenance: Battery and Patient Cable Replacement
1.
Press the internal access button to open the unit.
2.
Slide the battery release button in the direction of the arrow and lift the battery
out.
3.
Install a new battery and close the unit.
1.
Press the internal access button to open the unit.
2.
Press the connector release tabs and pull the connector loose.
3.
Pull the cable from the retaining tabs.
4.
Reassemble the cable by reversing the above steps.
MAC 5500 resting ECG analysis system
2020299-020
36A
37A
Revision E