System Errors

Problem
appears on the screen.
flashes intermittently.
appears on the screen.
The system does not power up when
operating from battery power.
The system shuts down when operating
from battery power.
"X" Lead disconnected message appears. Electrode(s) disconnected.
MODEM ERROR. The remote device is
not responding. Would you like to retry?
Cannot use the system because Device
Password does not work.
Revision E
Troubleshooting: System Errors
The following errors may occur while you are operating this system. You may be
required to perform some action.
If you perform the recommended actions and the condition still remains, contact
authorized service personnel. See "How to Reach Us" to find out how to contact GE
Medical Systems Information Technologies.
Cause
No battery is installed in the system.
The battery charge is low.
The writer door is open.
The battery is empty.
Battery is empty, or the Automatic Shutdown
feature is enabled.
Modem not connected. (If wireless option,
client bridge not connected.)
(Wireless option only) MAC 5500 is not
within range of access point.
Device Password has been changed or has
not been adequately communicated to the
staff.
NOTE
For information about troubleshooting the MobileLink Standard Security
option, see "MobileLink Installation & Troubleshooting Guide"
(PN 2020299-021).
For information about troubleshooting the MobileLink Ultra High Security
option, see "MobileLink UHS Installation & Troubleshooting Guide"
(PN 2020300-051).
MAC 5500 resting ECG analysis system
2020299-020
Solution
Install a battery and connect the system to
an AC wall outlet to charge the battery.
Connect the system to an AC wall outlet to
charge the battery.
Close the writer door.
Connect the system to an AC wall outlet to
charge the battery.
Connect the system to an AC wall outlet to
charge the battery, or power on the system.
Reconnect the electrode(s).
Connect and retry.
Relocate MAC 5500 to within range of
access point and retry transmission.
Override the Device Password prompt by
pressing the following keys at the same time:
i
63A, 64A, 65A