Guidance And Manufacturer's Declaration - Electromagnetic Immunity - GE MAC 5500 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 5500:
Table of Contents

Advertisement

Appendix C – Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
Immunity Test
Electrostatic discharge
(ESD)
± 6 kV contact
EN 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
±1 kV for input/output lines
EN 61000-4-4
Surge
± 1 kV differential mode
EN 61000-4-5
± 2 kV common mode
Voltage dips, short
<5% Ut (>95% dip in U
interruptions and
for 0.5 cycles
voltage variations on
40% Ut (60% dip in U
power supply input
for 5 cycles
lines
70% Ut (30% dip in U
EN 61000-4-11
for 25 cycles
<5% Ut (>95% dip in U
for 5 sec
Power frequency
(50/60 Hz) magnetic
3 A/m
field
EN 61000-4-8
C-4
The MAC 5500 is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or user to ensure that the MAC 5500 is
used in such an environment.
EN 60601 Test Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
±1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
<5% Ut (>95% dip in U
)
t
for 0.5 cycles
)
40% Ut (60% dip in U
t
for 5 cycles
)
70% Ut (30% dip in U
t
for 25 cycles
)
<5% Ut (>95% dip in U
t
for 5 sec
3 A/m
NOTE
Ut is the AC mains voltage prior to application of the test level.
MAC 5500 resting ECG analysis system
Compliance Level
)
t
)
t
)
t
)
t
2020299-020
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power should be that of a typical
commercial or hospital environment.
Mains power should be that of a typical
commercial or hospital environment.
Mains power should be that of a typical
commercial or hospital environment. If the
user of the MAC 5500 requires continued
operation during power mains interruptions,
it is recommended that the MAC 5500 be
powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should be
at levels characteristics of a typical location
in a typical commercial or hospital
environment.
Revision E