12-Port RF Cable Connections

Figure 10 on page 18
Set and how it should be configured to the E8364B/C 2-Port and N5230A/C PNA Network
Analyzer.
1. Connect the RF cables supplied with this option between the network analyzer and the
U3025AE10 Test Set. Torque each cable to 8 in-lb. Refer to
page
18.
Each end of the interconnect RF cables have a different length from the bend.
CAUTION
When connecting the RF Interconnect cables be sure that the longer end from
the bend is connected to the PNA.
CAUTION
Over torque will cause damage to the test set and may cause connectors to
spin or become loose.
Table 8
12-Port Cable Connection
RF Cables
From (PNA)
U3025-20044
SOURCE OUT
U3025-20044
CPLR THRU
U3025-20045
CPLR ARM
U3025-20045
RCVR A IN, B IN
User's Guide
illustrates the setup configuration of the U3025AE10 Multiport Test
To (Test Set)
SOURCE OUT
CPLR THRU
CPLR ARM
A IN, B IN
System Setup E8363B/C, E8364B/C, or N5230A/C
Table 8
and
Figure 10 on
17