Enabling Backup (Baup) - Omron Z4LB V2 Instruction Manual

Parallel beam linear sensor
Table of Contents

Advertisement

Key Settings
4-2-6

Enabling Backup (baUp)

Determine whether the forced-zero value is to be backed up or
not, i.e., to hold the value after the Sensor is turned OFF. If the
backup is not enabled, the value will be lost when the Sensor is
turned OFF.
Note If the backup is enabled, the forced-zero value will be written to
the internal EEPROM, which is a nonvolatile. The EEPROM has
a limit on the number of writing times. Therefore, if the forced-
zero value is changed frequently, disable the backup to increase
the life of the memory.
Procedure
scale
Measurement
width
1 Set the operating mode selector to RUN.
2 Press the TEACH Key for 3 s.
The Sensor will enter function selection mode.
3 Press the Up Key to go to the backup.
oUt
Discrimination
outputs
4 Press the SET Key.
The backup can now be set.
5 Press the Up Key to enable or disable the backup.
on
Enabled
6 Press the SET Key
The backup will be entered and the Sensor will return to
function selection mode.
7 Press the TEACH Key to return to RUN mode.
hys
Hysteresis
setting
baUp
Backup
Bank selection
off
Disabled
Section
4-2
hold
Hold
bank
45