64 | DRY 3.4 | Advanced Operation (key-operator mode)
Press the Confirm key to retry the calibration, or the Escape key to quit.
5. In case the film calibration has completed successfully, the following
message appears:
• The percentage for the 'Registration compensation' reflects the degree
• The percentage for the 'print head profile compensation' reflects the
Both values are important for the service technician.
6. Check the printed test films.
The test films must be similar to the images below.
Figure 1: Registration - TPH profile
2604A EN 20180322 1454
to which the system had to compensate for incorrect horizontal
positioning of the built-in densitometer (100% compensation =
maximal correction).
degree to which the system had to compensate for density variations in
the print head profile (100% compensation = maximal correction).
WARNING:
When the calibration process is interrupted, you will have to
start the calibration procedure again. Otherwise, the old
calibration parameters will be used.
WARNING:
Remark: the calibration film must be free of dust particles or
any other artifacts. If this is not the case, you must restart
the calibration procedure.