Nellcor OxiMax N-550 Service Manual page 43

Hide thumbs Also See for OxiMax N-550:
Table of Contents

Advertisement

Accessing Menu Items
Menu Item 1 (Trend Print)
N-550
default values that have been stored. Factory-set power-on default values are
listed in Table 4 on page 38.
1. Menu items can be accessed at any time by pressing the SpO
Limit and Pulse Rate Alarm Limit buttons simultaneously until the
option 1 screen is displayed.
2. Press the Adjust Up or Adjust Down button to select the menu item
number. Menu items 4, 5, 6, 7, 8, 11, 12, 13, and 14 have sub-options
within them that can be selected by first pressing the SatSeconds Alarm
Limit button, and then pressing the Adjust Up or Adjust Down button.
The sub-option being selected is displayed in the %SpO
Note: Service menu items numbered above 5 cannot be accessed if a sensor
is connected to the N-550.
3. Once adjustments have been made within a menu item, the SatSeconds
Alarm Limit button can be used to initiate the current selection. To save
the current settings as power-on default values, refer to the procedure
outlined in Menu Item 9 (Save Current Values as Power-On Defaults) on
page 37.
Note: If a period of 10 seconds passes with no button presses, the N-550
exits the menu item selected and return to normal monitoring, and no
changes will are made.
Trend data can be viewed (if connected to a PC), or a trend printout can be
made, if the SatSeconds Alarm Limit button is pressed when menu item 1 is
displayed. For more information about trend printouts, refer to the N-550
Operator's manual.
Audible Alarm Settings and Service Functions
Alarm