Nellcor OxiMax N-550 Service Manual page 36

Hide thumbs Also See for OxiMax N-550:
Table of Contents

Advertisement

Performance Verification
26
2. The N-550 pulse amplitude indicator will fill (10 bars) and stabilize and
illuminate 4 bars with each pulse beat.
3. The N-550 will display:
75 %SpO2 (test pass criteria is 73 to 77 %SpO
60 bpm (test pass criteria is 57 to 63 bpm)
alarm
pulse amplitude indicator - high level modulation
4. Perform Test #1: BPM on page 22. The pulse amplitude indicator should
indicate high level modulation.
5. Perform Test #2: SpO2 on page 23. The pulse amplitude indicator should
indicate high level modulation.
6. Press the SRC-MAX Light Level selection button. The SRC-MAX Light
Level
LED will light.
7. The N-550 pulse amplitude indicator will stabilize and illuminate 4 bars
with each pulse beat.
8. The N-550 will display:
75 %SpO2
60 bpm
alarm
low level modulation
)