Nellcor OxiMax N-550 Service Manual page 97

Hide thumbs Also See for OxiMax N-550:
Table of Contents

Advertisement

Safety Tests
Ground Integrity
Earth Leakage Current
Enclosure Leakage Current
N-550
100 milliohms or less
AC Line
AC Line
Polarity
Cord
Normal
Closed
Reversed
Closed
Normal
Open
Normal
Closed
AC Line
Neutral Line
Polarity
Cord
Normal
Closed
Normal
Closed
Normal
Open
Reversed
Closed
Reversed
Open
Reversed
Closed
-
* = AAMI/ANSI
ES1 does not include opening line conductor.
IEC 60601
Neutral
AAMI/
Line Cord
-
ANSI
ES1
Closed
µ
500
A
Closed
µ
500
A
Closed
µ
1000
A
Open
µ
1000
A
Power Line
Ground
IEC 60601
Cord
µ
Closed
100
µ
Open
500
Closed
µ
500
Closed
µ
100
Closed
µ
500
Open
µ
500
Specifications
-
1,
-
UL 2601
1
300 µA
300 µA
500 µA
500 µA
-
UL 2601
1,
-
1
AAMI/
-
ANSI
ES1
µ
A
100
A
µ
A
300
A*
µ
A
300
A
µ
A
100
A
µ
A
300
A*
µ
A
300
A
87