Test #4: Light
N-550
6. Press the SRC-MAX % Modulation selection button. The SRC-MAX %
Modulation
LED will light.
7. The N-550 pulse amplitude indicator will stabilize and illuminate 4 bars
with each pulse beat.
8. The N-550 will display:
•
75 %SpO2
•
60 bpm
•
alarm
•
pulse amplitude indicator - low level modulation
9. Perform Test #1: BPM on page 22. The pulse amplitude indicator should
indicate low level modulation.
10. Perform Test #2: SpO2 on page 23. The pulse amplitude indicator should
indicate low level modulation.
1. Press the SRC-MAX Light Level selection button. The SRC-MAX Light
Level
LED will light.
Performance Verification
25