Nellcor OxiMax N-550 Service Manual page 112

Hide thumbs Also See for OxiMax N-550:
Table of Contents

Advertisement

Technical Discussion
O
M
Technology
XI
AX
MP-506 Interface
102
The N-550 is designed to use Nellcor brand sensors containing O
technology. These sensors can be identified by their deep blue plug color. All
O
M
-compatible sensors contain a memory chip carrying information
XI
AX
about the sensor which the oximeter needs for correct operation, including the
sensor's calibration data, model type, troubleshooting codes, and error
detection data. This unique oximetry architecture enables several new features
with the N-550.
When an O
M
-compatible sensor is connected to the N-550, the N-550 will
XI
AX
first read the information in the sensor memory chip, check it to make sure
that there are no errors, and then load the data to begin monitoring. As the
N-550 reads the information, it flashes the sensor model number on its
display. This process only takes a couple of seconds. Once the reading process
is complete, the sensor model number will stop flashing on the display, and
then the N-550 will begin monitoring. The sensor model number disappears
once the N-550 starts tracking the patient's SpO
Monitors containing O
XI
data contained in the sensor in calculating the patient's SpO
calibration in the sensor, rather than the N-550, Nellcor is able to improve the
published accuracy of many sensors, because the calibration coefficients can
be tailored to each sensor. Consult the accuracy card included with the N-550
for specific accuracy information for the N-550 with different Nellcor
sensors.
The N-550 uses the information in the O
troubleshooting messages for the clinician. The sensor contains coding that
tells the N-550 what kind of sensor is being used. When deciding what
messages to display, the N-550 takes into account the sensor type and
recommended patient site for that model. The N-550 O
has an intelligent troubleshooting system.
The MP-506 interfaces with the N-550 through two connections. Figure 45
illustrates the interconnection, and Table 15 and Table 16 identify the signals.
2
M
technology, including the N-550, use calibration
AX
M
-compatible sensor to tailor
XI
AX
M
XI
AX
and pulse rate.
. By having the
2
M
system therefore
XI
AX

Hide quick links:

Advertisement

Table of Contents
loading