Nellcor OxiMax N-550 Service Manual page 61

Hide thumbs Also See for OxiMax N-550:
Table of Contents

Advertisement

N-550
Table 11: Error Codes (Continued)
Error
Meaning
Code
258
SpO2 back end reports packet length error
259
SpO2 back end reports message length error
260
SpO2 back end reports packet contains unsupported Key
261
SpO2 back end reports packet CRC error
262
SpO2 back end reports end of packet missing
263
SpO2 back end reports packet contains undefined key
264
SpO2 back end reports corrupted variable
265
SpO2 back end reports memory overflow
266
SpO2 back end reports bad pointer
267
SpO2 back end reports parameter value out
268
SpO2 back end reports reset detected
269
SpO2 back end reports unexpected value
270
SpO2 back end reports time
271
SpO2 back end reports not ready/not initialized
272
SpO2 back end reports double fault
273
SpO2 back end reports date out
274
SpO2 back end reports incompatible software version
275
SpO2 back end reports incorrect registration number
276
SpO2 back end reports sensor read failure
277
SpO2 back end reports sensor signature verification fails
278
SpO2 back end reports warmed sensor temperature set point failure
279
SpO2 back end reports warmed sensor/SpO2 front end incompatible
280
SpO2 back end reports does not support feature required by sensor
281
SpO2 back end reports overflow/underflow
282
SpO2 back end reports sensor activation failure
512
General failure of UIF Module generic post
512
Dead battery/Missing battery
514
Real time clock is non
517
Serial clock line is not toggling or is toggling at an incorrect rate
-
-
of
range
-
out
-
-
of
range error
-
operational