Introduction - Nellcor OxiMax N-550 Service Manual

Hide thumbs Also See for OxiMax N-550:
Table of Contents

Advertisement

Performance Verification

Introduction

20
1 — DOC-10 Cable Connector
2 — Red LED Drive Indicator
3 — Not Used For N-550
4 — Not Used For N-550
5 — Battery Low Indicator
The SRC-MAX functional tester allows qualified technicians to functionally
test Nellcor O
M
technology-based pulse oximeters and OEM O
XI
AX
technology-based monitors. The technician must perform the test setup
procedure before performing tests 1 through 4. The following is a brief
description of each test:
Test Setup — This procedure establishes the baseline for all the other
tests. The Test Setup procedure must be performed before performing
any or all of the SRC-MAX tests.
Test 1: BPM — This procedure simulates an O
attached to a patient indicating 60 bpm and 200 bpm. The test setup
procedure sets up Test 1 for 60 bpm.
Test 2: SpO
— This procedure simulates an O
2
attached to a patient, indicating 75 percent blood oxygen saturation
and 90 percent blood oxygen saturation. The test setup procedure sets
up Test 2 for 75 percent blood oxygen saturation.
6 — % Modulation Select Button
7 — % SpO
Select Button
2
8 — Light Level Selection Button
9 — Pulse Rate Selection Button
10 — Infrared LED Drive Indicator
M
XI
AX
M
XI
AX
M
XI
AX
sensor
sensor